Operating your VS-2501

For best results make sure your unit is on a level and stable surface

Zeroing.

Before weighing the Weight Display needs to be showing 0.0. If not a simple press of th
e                                                                                                                                                                                                             
 

 

Key will reset the unit to zero. 

  
Straight Weighing

Use care when placing the items to be weighed not drop heavy items on the unit. Once the 
subject is one the scale the weight will be displayed. Upon removal the weight display will 
return to 0.0

Weighing With Tare.

Tare weighing provides a weight by fi rst subtracting the weight of the object not needed in 
the fi nal result. 
 

1. Place the object onto the platform

 2. 

Press 

 

   * Tare enunciator with display*

 

3. The Weight display will return to zero

 

4. Place the actual weight including the tared container onto the platform. 

 

    The Weight displayed is the actual weight of the object not including the container.

 

5. To remove Tare value press       with no weight on Platform

Accumulation Operation

With the unit at 0.0 load the fi rst weight and wait until stable. Press       activates the accu-
mulation mode. Total enunciator with illuminate, to show accumulation mode is active. 
 

 

Display: n 001 First weight captured

 

 

Remove First weight and add second weight

  Press 
 

 

Display: n 002 Second Weight Captured ( maximum 999 weights )

To show Total
 

 

Press      and         at the same time, Display show N xxx, where xxx is the  

 

 

number of weighment, and then shows total accumulated weight. To continue  

 

 

accumulation Press       for 2 seconds and then Press  

 

Visit www.scaleman.com to purchase your WeighSouth VS2501

Visit www.balances.com to purchase your WeighSouth VS2501

Summary of Contents for VS-2501

Page 1: ...VS 2501 Veterinary Scale Operations Manual Visit www scaleman com to purchase your WeighSouth VS2501 Visit www balances com to purchase your WeighSouth VS2501...

Page 2: ...ith the scale The indicator is static and sensitive equipment Remove power during electrical servicing Internal components touched by hand will void warranty Handle the scale with care Do not put exce...

Page 3: ...ully loaded scale Assembling 1 With the box opened you will notice the large platform upside down While in the box screw in the 4 adjustable feet into each corner of the base assembly Make sure that e...

Page 4: ...Battery light Battery Charged Battery Operation Your VS 2501 is equipped with an internal rechargeable battery The unit will automati cally switch to battery mode when the power supply is disconnected...

Page 5: ...es Tared Weight Values Gross Weight is shown Indicates a Tare value is stored Weighing unit is pounds Battery Indicates check weighing function Weight is Zero and Stable The weighing data is stable Gr...

Page 6: ...zero 4 Place the actual weight including the tared container onto the platform The Weight displayed is the actual weight of the object not including the container 5 To remove Tare value press with no...

Page 7: ...mal by placing the animal on the platform Once the animal becomes stable within tolerances the VS 2501 will calculate the average from various readings Once the average is set the VS 2501 will lock in...

Page 8: ...with each C Function having different settings Entering Set Up Menu Press C01 Navigating Menu Structure 1 C Function C Sub Menu Enters C Function Exits C Sub Menu Exits and Saves C Functions Only Exc...

Page 9: ...g menu C01 C39 step Method of operation display Remark 1 C01 After you enter calibration mode it display C01 2 press C01 1 Weight unit option 1 kg 2 lb 3 press press press or C02 C02 0 C02 2 Set decim...

Page 10: ...lease choose 1 and ensure scale is empty and stable light is on Ensure zero calibration countdown Till show 0 00 example for two decimal point 7 press press press or press press or press C06 C06 0 C06...

Page 11: ...ting Item parameters setting and instruction warning tone C08 warning tone Options 0 close warning tone 1 open warning tone Automatic power off C09 Automatic power off option 0 close auto power off 10...

Page 12: ...2 1 kg lb conversion is ok C13 Upper limit alarm value Upper lower limit alarm C14 Lower limit alarm value You can set it within the max capacity limit Inner Code display C15 Check inner code enter C1...

Page 13: ...pacity 10 10 max capacity 20 20 max capacity 100 100 max capacity Zero range C21 Initial zero range option 0 no initial zero setting 1 1 max capacity 2 1 max capacity 5 1 max capacity 10 1 max capacit...

Page 14: ...ge 01d 99d remark d division Negative display C25 Negative display range Option 0 9d 10 10 max capacity 20 20 max capacity 50 50 max capacity 100 100 max capacity C26 Standstill time Option 0 quick 1...

Page 15: ...er more stable if the loaded weight shake not too much The setting is less than 3 C29 Noise filter option 0 close noise filter 1 1 digital filter strength 2 2 digital filter strength 3 3 digital filte...

Page 16: ...unication C35 C35 0 99 signal Gravity of calibration location C36 C36 9 7000 9 9999 Gravity of destination C37 C37 9 7000 9 9999 Version No C38 Preserved menu C39 Visit www scaleman com to purchase yo...

Page 17: ...0 0 One Decimal Point 2 0 00 Two Decimal Point Scale Graduation C03 C03 1 0 0 1 0 1 2 0 2 5 0 5 Scale Maximum Capacity C04 0 Scale Zero Calibration C05 C05 0 0 No Zero Calibration 1 Zero Calibrate Sca...

Page 18: ...put and output resistance ERR1 During calibration not input the weights or the weight is overload 1 Input the correct weights ERR2 During calibration the input signal is nega tive The calibration weig...

Page 19: ...r replacement or credit at The Manufacturer option Any product repaired or replaced under the foregoing warranty shall be shipped back to customer freight and insurance prepaid Except as provided abov...

Reviews: