No�
Fault
Can be recognized by
Possible cause
Remedial measures
4
Segments of the clamping
jaws do not separate widely
enough
Connector is seated too
loosely, it sits unsteadily
on the test piece
Wear
First check the test piece. If this is
defective, replace the test piece. If it is
in good condition, send the
connector to WEH for maintenance
Connector is not
correctly connected
Repeat the connecting procedure
5
Connector wobbles despite
a correct connection to the
test piece
Incorrect connection
geometry at test piece
Wear on geometry of the
test piece’s connection
parts
Replace test piece
Wear on the connector’s
clamping jaws
Return the connector to WEH
for maintenance
6
Clamping jaws deformed
The sliding sleeve can
no longer be connected
Excessive mechanical
stress due to
inappropriate handling
Return the connector to WEH
for maintenance
7
Clamping jaws do not close
The sliding sleeve can
no longer be connected
Wrong test piece
Use correct test piece
If you should encounter any other problems, please contact WEH or your responsible distributor.