Waygate Technologies DM5E Basic Quick Start Manual Download Page 6

Technical specifications

DM5E family

Instrument specifications

Operating principle

Ultrasonic, pulse echo measurement method

Measuring range

Depends on probe and material, 0.60 mm to 508 mm (0.025” to 20.00”) in IP to 1st BW measurement mode,  

2.00 mm to 127.0 mm (0.079” to 5.00”) in dual-multi measurement mode, the coating thickness range shall  

be 0.3 mm to 2.50 mm (0” to 0.098”).

Measuring  

resolution

0.01 mm default - selectable 0.01, 0.1 mm (0.001” default - selectable 0.001”, 0.01”)

Material velocity 

range

0.508 to 18.699 mm/msec (0.0200” to 0.7362”/msec)

Material velocity 

resolution

1 m/s (0.0001”/msec)

Units

Inch or millimeter

Calibration

One-point cal, on-block and off-block, two-point cal

Pulser

Excitation Pulse

Spike pulser

Voltage

120 V into 50 ohm load, using 20 MHz oscilloscope

Receiver

Bandwidth

500 KHz to 12 MHz @ -3 dB

Gain

Automatic gain control

Display type

High-resolution graphical LCD, 64 x 128 pixels, 53.0 mm x 27.0 mm with backlight and adjustable contrast

Update rate

4 Hz or 8 Hz, user selectable, 24 Hz scan mode capture rate

Thickness value 

display

NORMAL view mode

5 Digit, 10.6 mm (0.4”) high

B-SCAN view mode

5 Digit, 2.55 mm (0.1”) high

Display of last 

reading

Solid filled or hollow digits indicate coupled or uncoupled condition

Setups

9 standard setups for probes

Alarm settings

Minimum and maximum alarms, range of 0.25 mm to 508 mm, 0. (0.010” to 20.00”) 

reading alternates between solid and hollow when alarms are enabled and violated

Power  

requirements

2 “AA” size batteries

Battery life/

operating time

Approximately 60 hours

Instrument shut-off

Selectable ALWAYS ON or AUTO OFF after 5, 10, 15, 30 minutes of inactivity

Language

Selectable English, German, French, Spanish, Italian, Russian, Japanese and Chinese

I/O connectors

Transducer

Dual Lemo 00 (coax)

Mini-USB

Mini USB to PC

Temperature

Operating

-10°C to +50°C (+10°F to +120°F)

Storage

-20°C to +60°C (-10°F to +140°F)

Weight

223 g (0.597 lb) including batteries

Size

138 mm x 32 mm x 75 mm

Shock

IEC 68-2-27 Ea, as per Mil Std 810C Method 516.2 Procedure I with a 15g 11ms impulse half sinusoidal wave 

applied 6 times per axis

Sealing

IEC529 / IP54, dust proof/dripping water proof as per IEC 529 specifications for IP54 classification

Summary of Contents for DM5E Basic

Page 1: ...DM5E corrosion thickness gauges A range of high performance reliable and easy to use instruments...

Page 2: ...er previous corrosion thickness gauges in terms of better thickness measurement stability and repeatability at normal and elevated temperatures It has been designed for operation in the harshest worki...

Page 3: ...and nominal thickness Choose from three levels of functionality DM5E Basic DM5E or DM5E DL The DM5E The DM5E incorporates all the features of the DM5E Basic but also offers the DUAL MULTI operating m...

Page 4: ...adings in files All calibration is menu driven and the operator is guided through every step There is a built in calibration reminder which can be set to remind the user to calibrate after a specified...

Page 5: ...om the fact that the maximum thickness measured is displayed DIFF RR Compares the measured thickness with a user specified nominal thickness The dimensional difference between the two values is displa...

Page 6: ...selectable 24 Hz scan mode capture rate Thickness value display NORMAL view mode 5 Digit 10 6 mm 0 4 high B SCAN view mode 5 Digit 2 55 mm 0 1 high Display of last reading Solid filled or hollow digit...

Page 7: ...Model DA590 Model FH2E Frequency 5 MHz 2 MHz 7 5 MHz 5 MHz 7 5 MHz Probe style Standard Standard Fingertip High temperature Fingertip Operating temperature range continuous 20 to 70 C 20 to 70 C 20 to...

Page 8: ...ce or obligation Contact your BHI representative for the most current information The Baker Hughes logo is a trade mark of Baker Hughes The BH Monogram is a trademark of Baker Hughes JB170718 04 2020...

Reviews: