Wavecom 89 User Manual Download Page 12

 

12 

Specifications 

Function 

Range 

Resolutio
n  

Accuracy 

DC Voltage 

 

600mV 

0.1mV 

(0.09% r 2 digits) 

6V 

0.001V 

60V 

0.01V 

600V 

0.1V 

1000V 

1V 

(0.15% r 2 digits) 

AC Voltage 

 

 

50 to 60Hz 

40Hz to 1KHz 

6V 

0.001V 

(1.0% r 3 

dgts) 

(2.0% r 3 dgts) 

60V 

0.01V 

600V 

0.1V 

1000V 

1V 

(1.2% r 3 

dgts) 

(2.5% r 3 dgts) 

 

All AC voltage ranges are specified from 5% of range to 100% of range 

DC Current  600 A 

0.1 A 

(1.0% r 3 digits) 

6000 A 

1 A 

60mA 

0.01mA 

600mA 

0.1mA 

6A 

0.001A 

(1.5% r 3 digits) 

10A 

0.01A 

(20A: 30 sec max with reduced accuracy) 

AC Current 

 

 

40Hz to 1kHz 

600 A 

0.1 A 

(1.5% r 3 digits) 

6000 A 

1 A 

60mA 

0.01mA 

600mA 

0.1mA 

 

6A 

0.001A 

(2.0% r 3 digits) 

10A 

0.01A 

(20A: 30 sec max with reduced accuracy) 

 

All AC voltage ranges are specified from 5% of range to 100% of range 

NOTE:

 Accuracy is stated at 65

o

F to 83

o

F (18

o

C to 28

o

C) and less than 75% RH. 

Summary of Contents for 89

Page 1: ...A VAC VDC MAX MIN TEMP CAP TEMP True RMS Industrial Multimeter MODEL WAVECOM 89 Wavecom Instruments Pty Ltd 257 Grange Road FINDON SA 5023 P 61 08 8243 3500 F 61 08 8243 3501 sales wavecom com au www...

Page 2: ...acent to one or more terminals identifies them as being associated with ranges that may in normal use be subjected to particularly hazardous voltages For maximum safety the meter and its test leads sh...

Page 3: ...EXTREME CAUTION when working with high voltages 3 DO NOT measure voltage if the voltage on the COM input jack exceeds 600V above earth ground 4 NEVER connect the meter leads across a voltage source w...

Page 4: ...e on rear of unit Note Input Jacks for test probes are a firm fit to prevent dislocation Take care to fully insert banana plug into jack Symbols and Annunciators Auto power off Continuity Diode test B...

Page 5: ...Risk of Electrocution The probe tips may not be long enough to contact the live parts inside some 240V outlets for appliances because the contacts are recessed deep in the outlets As a result the read...

Page 6: ...FREQUENCY DUTY CYCLE MEASUREMENTS CAUTION Do not make 20A current measurements for longer than 30 seconds Exceeding 30 seconds may cause damage to the meter and or the test leads 1 Insert the black te...

Page 7: ...continuity on circuits or wires that have voltage on them 1 Set the function switch to the CAP position 2 Insert the black lead banana plug into the negative COM jack Insert the red test lead banana...

Page 8: ...ve the batteries and unplug the line cords 1 Set the rotary function switch to the CAP position 2 Insert the black test lead banana plug into the negative COM jack Insert the red test lead banana plug...

Page 9: ...the minimum reading and will update only when a new min occurs 3 To exit MAX MIN mode press and hold the MAX MIN key for 2 seconds RELATIVE MODE The relative measurement feature allows you to make mea...

Page 10: ...chemicals cleaning solvents or detergents 5 USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE AND TYPE Remove old or weak batteries so they do not leak and damage the unit 6 IF THE METER IS TO BE STOR...

Page 11: ...the battery 4 Remove the six A screws securing the rear cover 5 Gently remove the old fuse and install the new fuse into the holder 6 Always use a fuse of the proper size and value 0 8A 1000V fast blo...

Page 12: ...ranges are specified from 5 of range to 100 of range DC Current 600 A 0 1 A 1 0 reading 3 digits 6000 A 1 A 60mA 0 01mA 600mA 0 1mA 6A 0 001A 1 5 reading 3 digits 10A 0 01A 20A 30 sec max with reduced...

Page 13: ...999MHz 0 001MHz 40MHz 01MHz Sensitivity 0 8V rms min 20 to 80 duty cycle and 100kHz 5Vrms min 20 to 80 duty cycle and 100kHz Frequency electrical 10 00 400Hz 0 01Hz 0 5 reading Sensitivity 15Vrms Dut...

Page 14: ...is displayed Auto Power Off 15 minutes approximately with disable feature Polarity Automatic no indication for positive Minus sign for negative Measurement Rate 2 times per second nominal Low Battery...

Page 15: ...15...

Page 16: ...Road FINDON SA 5023 Ph 61 08 8243 3500 Fax 61 08 8243 3501 Email sales wavecom com au Web www wavecom com au Melbourne 772A Station Street BOX HILL VIC 3128 Ph 61 03 9897 4711 Fax 61 03 9897 4766 Ema...

Reviews: