Watson-Marlow 323S User Manual Download Page 44

Watson-Marlow 323E, 323S, 323U and 323Du User Manual

44

The tube clamps can accommodate various tube diameters by pushing in, or pulling
out, the grip bars within the clamp. Set the clamps to apply the minimum necessary
pressure to the tubing.

Restart the pump. Free the downstream clamp for a short time, while the pump is
running, so that the tube can find its natural length. Please keep fingers clear of the
moving rotor. Shut and lock the guard after adjusting the tube.

When using Marprene tubing

Re-tension new tube after the first 30 minutes of running. Stop the pump and release
the tube clamp at the outlet of the pump. Pull any slack tube from the pumphead and
re-clamp the tube. Re-start the pump. This will correct the normal stretching that oc-
curs with new Marprene tube. The correct tension is essential for good tube life.

23.14 501RL and 501RL2 rotor settings

The 501RL and 501RL2 pumpheads are factory set to give optimum tube life with
Watson-Marlow tubing. We recommend that the rotors are not adjusted or that other
types of tubing are not used.

If the rotor needs re-aligning we recommend the rotor is returned to Watson-Marlow
for correct adjustment. Or contact our technical department for further information.

Check moving parts of the rotor from time to time for freedom of movement. Lubri-
cate pivot points and rollers occasionally with Teflon lubricating oil.

Summary of Contents for 323S

Page 1: ...ic control wiring using the 720N module 26 18 Care and maintenance 28 19 Troubleshooting 28 19 1 Error messages 29 20 Drive maintenance 30 21 Drive part numbers 30 22 Drive spares 31 23 Pumpheads 32 2...

Page 2: ...not be put into service until the relevant machin ery has been declared in conformity with the Machinery Directive 2006 42 EC Responsible person Christopher Gadsden Managing Director Watson Marlow Lim...

Page 3: ...itions of and specific exceptions to the above warranty are Conditions Products must be returned by pre arrangement carriage paid to Watson Mar low or a Watson Marlow approved service centre All repai...

Page 4: ...323U and 323Du pumps are supplied as Dedicated 323E 323S 323U or 323Du pump drive unit fitted with one or more 313 or 314 pumpheads see 8 Pump specifications The designated mains power lead for your...

Page 5: ...turned to Watson Marlow or its distributor A certificate included at the rear of these operating instructions or signed statement must be attached to the outside of the shipping carton This certificat...

Page 6: ...first roller leaves the track the second continues to advance expelling the packet of fluid through the pump s discharge port At the same time a new partial vacuum is created behind the second roller...

Page 7: ...t under the terms of the EU Waste Electrical and Electronic Equipment WEEE Di rective Fundamental work with regard to lifting trans portation installation starting up maintenance and repair should be...

Page 8: ...d only for its intended purpose The pump must be accessible at all times to facilitate operation and maintenance Access points must not be ob structed or blocked The pump s mains plug is the disconnec...

Page 9: ...control Backlight Beeper Manual screen rpm 15 400 rpm 27 1 speed control 3 400 rpm 133 1 speed control 1 5 220 rpm 147 1 speed control 5 rpm scrolling increment 1 rpm scrolling increment Auto restart...

Page 10: ...t ted with a short nosed gearbox offers 15 400 rpm and can be fitted with a 313 or 314 pumphead The 323S 323U and 323Du are available with a choice of gearboxes a short nosed gearbox which offers 3 40...

Page 11: ...against water projected from a nozzle against the equipment enclosure from any direction There must be no harmful effect water jet 12 Indoor use to provide a degree of protection against dust falling...

Page 12: ...A Full load current 0 43A at 230V 0 86A at 115V Eprom version Accessible through pump software Enclosure rating IP31 to BS EN 60529 Equivalent to NEMA 2 Suitable for indoor use Pro tected against drip...

Page 13: ...2 Class A called by BS EN 61000 6 4 Electrostatic discharge BS EN 61000 4 2 Radiated RF immunity BS EN 61000 4 3 A1 and A2 called by BS EN 61000 6 2 Fast transient burst BS EN 61000 4 4 A1 and A2 Leve...

Page 14: ...S 323U and 323Du User Manual 14 8 2 Dimensions in millimetres 323E D 323S D 323U D 323Du D 323S MC 323U MC 323Du MC 323S RL 323U RL 323Du RL Drive only 313 501RL 323 4 2kg 9lb 4oz 4 5kg 9lb 14oz 5 5kg...

Page 15: ...e local emergency stop device is fitted into the mains supply to the pump Valves Peristaltic pumps are self priming and self sealing against backflow No valves are required in inlet or discharge lines...

Page 16: ...p is running Do ensure that on longer tube runs at least 1m of smooth bore flexible tubing is con nected to the inlet and discharge port of the pumphead to help to minimize impulse losses and pulsatio...

Page 17: ...mains borne noise The voltage selector is mounted in the switchplate at the rear of the pump Set the voltage selector to 115V for 100 120V 50 60Hz supplies or 230V for 200 240V 50 60Hz supplies Al wa...

Page 18: ...has been made to a suitable power supply Ensure that the recommendations in section 9 Good pump installation practice are followed Check the position of the voltage selector switch Check the mains po...

Page 19: ...Mains power must be available to the pump to engage auto restart Stop the pump Turn off the mains power switch at the rear of the pump Hold down the START key and turn the mains power switch on The s...

Page 20: ...irection is shown by the rotation symbol The di rection may be changed while the pump is stopped or running Start the pump with the START key The rotation symbol will move to confirm that the pump is...

Page 21: ...o lock the keypad The padlock symbol will show and only the START and STOP key will function The keypad may also be locked while the pump is stopped Hold down the STOP key for more than 2 seconds The...

Page 22: ...ispensed If the quantity is correct press START to repeat the dose If the initial dose is larger than required use the DOWN key to reduce the volume showing on the display This will reduce the next do...

Page 23: ...ump speed Check that the pump is ready to run before selecting automatic operation Remote control signals may start the pump without warning Press the MODE key to select automatic operation The pump w...

Page 24: ...rror message E21 Over signal See 19 Error messages The remote stop start input works with both manual and analogue control modes The remote direction input works only with analogue control mode Never...

Page 25: ...remote stop start switch If you invert the operation of the remote stop start switch you must connect a wire link from pin 7 to 17 to be able to start the pump from the keypad This diagram shows the c...

Page 26: ...17 2 RS232 serial link The RS232 interface will give basic pump control by serial link to the 9 way D con nector at the rear of the pump To select RS232 serial control repeat MODE key press until dig...

Page 27: ...unction 1SPxxx Set pump speed to xxx 1RC Reverse direction 1SI Increase speed by 1rpm 1RR Set clockwise direction 1SD Reduce speed by 1rpm 1RL Set counter clockwise direction 1GO Start pump 1RS Show a...

Page 28: ...n Marlow or its appointed agents or distributors for service 19 Troubleshooting If the pump display remains blank when the pump is on make the following checks Check the position of the voltage select...

Page 29: ...upport 15 Over current Stop pump immediately Check system Power OFF ON may reset Or seek support 16 Over voltage Stop pump immediately Check mains voltage se lector switch Check supply Power OFF ON ma...

Page 30: ...00U 323S 220 N A UK 036 3134 00U 323S 400 N A UK 036 3142 00U 323U 220 N A UK 036 3144 00U 323U 400 N A UK 036 3152 00U 323Du 220 N A UK 036 3154 00U 323Du 400 N A UK Complete pump assemblies Part nu...

Page 31: ...al 31 E S U Du 22 Drive spares Spare Description 1 MNA2042A 313 and MC bayo net mounting plate 400 rpm models only 2 MN2056M E and S interface card cover 3 MN2094T U interface card cover 4 MN2095T Du...

Page 32: ...and is designed to provide higher flow rates The 314D pumphead has four rollers to provide greater pumping precision with less pulsation in the flow Both designs are available for 1 6mm and 2 4mm wal...

Page 33: ...n the pumphead clockwise until it locks into an upright position Removal Push the locking lever back and turn the pumphead anti clockwise until it is free from the mounting plate Tube loading Switch o...

Page 34: ...to resettle naturally across the rollers Re clamp the tube Re start the pump This will correct the normal stretching that occurs with new Marprene tube The correct tension is essential for good tube...

Page 35: ...rpm 1 5 220 rpm 0 45 12 0 09 12 0 05 6 6 1 1 28 0 21 28 0 11 15 4 1 110 0 81 110 0 41 59 15 400 3 0 400 1 5 220 33 880 6 6 880 3 3 480 54 1400 11 1400 5 4 790 75 2000 15 2000 7 5 1100 Flow rates 314D...

Page 36: ...Tygon Neoprene Fluorel 0 2 bar bore mm 0 5 0 8 1 6 3 2 4 8 6 4 8 0 in 1 50 1 32 1 16 1 8 3 16 1 14 5 16 112 13 14 16 25 17 18 220 400 rpm 6 6 4 2 2 1 1 313D 314D STA PURE CHEM SURE 0 5 2 bar bore mm 1...

Page 37: ...903 0080 016 965 0080 016 913 A080 016 1 6mm tube mm in PVC Fluorel Neoprene STA PURE 0 8 1 32 13 920 0008 016 1 6 1 16 14 950 0016 016 970 0016 016 920 0016 016 960 A016 016 3 2 1 8 16 950 0032 016 9...

Page 38: ...design Cassettes locate quickly with a single cam lever action that also controls roller pressure against the tube Extension pumpheads may be added up to 10 channels All are bayonet mounted for easy...

Page 39: ...the cassette legs Slide the tubing evenly down into the slots Do not twist the tube or use unnecessary force Slip the collars past the retaining fingers and ensure the tubing has reached the bottom of...

Page 40: ...d spares Spare Description 1 033 6453 000 314MC four roller five channel pumphead 1 033 6454 000 314MCX four roller five channel extension pumphead 1 033 6853 000 318MC eight roller five channel pumph...

Page 41: ...35 1 8 13 yellow yellow 1 42mm 0 055in 0 46 2 3 17 yellow blue 1 52mm 0 060in 0 52 2 6 19 blue blue 1 65mm 0 065in 0 60 3 0 22 green green 1 85mm 0 070in 0 76 3 8 28 purple purple 2 05mm 0 080in 0 90...

Page 42: ...0 050in 979 0129 000 981 0129 000 983 0129 000 yellow yellow 1 42mm 0 055in 979 0142 000 981 0142 000 983 0142 000 yellow blue 1 52mm 0 060in 979 0152 000 981 0152 000 983 0152 000 blue blue 1 65mm 0...

Page 43: ...a drive pin inside the rotor collet to engage with the end of the drive shaft To keep this pin correctly engaged on the drive shaft we recommend that the collet remains on the shaft while the rotor i...

Page 44: ...e clamp at the outlet of the pump Pull any slack tube from the pumphead and re clamp the tube Re start the pump This will correct the normal stretching that oc curs with new Marprene tube The correct...

Page 45: ...ad 053 0001 L20 501RL2 complete pumphead 1 SG001 SG002 Springs for 501RL blue Springs for 501RL2 red 2 MN0012T Follower roller 3 MN0011T Main roller 4 MNA0143A 501RL rotor assembly 5 CL0656T Collet 6...

Page 46: ...016 965 0016 016 913 A016 016 3 2 1 8 16 902 0032 016 903 0032 016 965 0032 016 913 A032 016 4 8 3 16 25 902 0048 016 903 0048 016 965 0048 016 913 A048 016 6 4 1 4 17 902 0064 016 903 0064 016 965 0...

Page 47: ...a Pure and Chem Sure are trademarks of W L Gore and Associates 25 Warning not to use pumps in patient connected applications Warning These products are not designed for use in and should not be used f...

Page 48: ...contamination certificate for each product You are responsible for cleaning and decontaminating the product s before return Your name Company Address Postcode zip Country Telephone Fax Product type Se...

Reviews: