background image

6

XSelect 

XP

 and XSelect Premier 2.5 µm HSS Columns

[ CARE AND USE MANUAL ]

b. Normal-phase Conditions

The HSS Cyano column can be used for both reversed-phase 
separations as well as normal-phase separations. The column 
is originally shipped in acetonitrile and is ready to use for 
reversed-phase conditions.

If you intend to use the column for normal-phase 
applications, you will need to condition the column 
with the following procedure:

1.  Flush the column with a minimum of 20 column 

volumes of 100% methanol using a low flow rate to 
avoid overpressuring the LC system. Refer to Table 1 
for minimum solvent volume.

2.  Flush the column with a minimum of 20 column 

volumes of 100% isopropanol using a low flow rate 
to avoid overpressuring the LC system. Refer to 
Table 1 for the minimum solvent volume.

3.  Flush the column with a minimum of 20 column volumes 

of 100% dichloromethane using a low flow rate to avoid 
overpressuring the LC system. Refer to Table 1 for the 
minimum solvent volume.

4.  Flush the column with the intended mobile-phase 

conditions until a stable baseline is achieved.

c. Storage

For periods longer than four days at room temperature, 
store reversed-phase XSelect HSS Columns in 100% 
acetonitrile. For elevated temperature applications, store 
immediately after use in 100% acetonitrile for the best 
column lifetime. Do not store columns in buffered eluents. 
If the mobile phase contained a buffer salt, flush the column 
with 10-column volumes of HPLC-grade water (see Table 
1 for common column volumes) and replace with 100% 
acetonitrile for storage. Failure to perform this intermediate 
step could result in precipitation of the buffer salt in the 
column when 100% acetonitrile is introduced. Completely 
seal column to avoid evaporation and drying out of the bed.

Note: If a column has been run with a mobile phase that contains 
formate (e.g., ammonium formate, formic acid, etc.) and is then 
flushed with 100% acetonitrile, slightly longer equilibration times 
may be necessary when the column is re-installed and run again 
with a formate-containing mobile phase.

d. Post Normal-phase Use

For rapid equilibration upon startup, it is recommended that 
you store the HSS Cyano column in the mobile phase that is 
commonly used for your normal-phase separation. Completely 
seal the column to avoid evaporation and drying out of the bed.

IV. eCORD INTELLIGENT CHIP TECHNOLOGY

a. Introduction

The eCord Intelligent Chip provides the history of a column’s 
performance throughout its lifetime. The eCord will be 
permanently attached to the column to assure that the 
column’s performance history is maintained in the event that 
the column is moved from one instrument to another.

At the time of manufacture, tracking and quality control 
information will be downloaded to the eCord. Storing this 
information on the chip will eliminate the need for a paper 
Certificate of Analysis. Once the user installs the column, 
the software will automatically download key parameters 
into a column history file stored on the chip. In this manual, 
we explain how the eCord will provide a solution for easily 
tracking the history of the columns, reduce the frustration of 
paperwork trails, and give customers the reassurance that a 
well-performing column is installed onto their instruments.

b. Installation

Install the column into the column heater. Plug the eCord 
into the side of the column heater. Once the eCord is inserted 
into the column heater the identification and overall column 
usage information will be available allowing the user to access 
column information on their desktop.

Figure 3. eCord Intelligent Chip.

eCord Intelligent Chip

Figure 4. eCord inserted into side of column heater. 

eCord inserted into 

side of column heater

Summary of Contents for XSelect XP

Page 1: ...ra pure reagents Each batch of XSelect HSS material is tested chromatographically with acidic basic and neutral analytes and the results are held to narrow specification ranges to assure excellent rep...

Page 2: ...ents entry of air into the detection system 4 Gradually increase the flow rate as described in Step 2 5 Once a steady backpressure and baseline have been achieved proceed to the next section Note If m...

Page 3: ...d to the inlet of the column For more information on VanGuard products visit www waters com vanguard g Replacing the VanGuard FIT Cartridge The VanGuard FIT Cartridge is designed specifically for XSel...

Page 4: ...nd mobile phases are miscible in order to avoid sample and or buffer precipitation Filter sample with 0 2 m membranes to remove particulates If the sample is dissolved in a solvent that contains an or...

Page 5: ...d cleaning call your local Waters office for additional support Table 3 Recommended pH and Temperature Limits for XSelect HSS Columns Column Particle Size Pore Diameter Surface Area pH Limits Temperat...

Page 6: ...out of the bed Note If a column has been run with a mobile phase that contains formate e g ammonium formate formic acid etc and is then flushed with 100 acetonitrile slightly longer equilibration tim...

Page 7: ...ncludes mobile phases running conditions and analytes used to test the columns In addition the QC results and acceptance is placed onto the column Figure 5 The eCord chip provides the user with an ove...

Page 8: ...nal HPLC column and therefore mobile phases last much longer To reduce the chances of mobile phase contamination or degradation only prepare what you need for analysis or store excess bulk quantities...

Reviews: