Ion optics
1-13
Ion optics
The mass spectrometer’s ion optics operate as follows:
1. Samples from the LC or Intellistart fluidics system are introduced at
atmospheric pressure into the ionization source.
2. The ions pass through the sample cone into the vacuum system.
3. The ions pass through the transfer optics (the ion guide) to the first
quadrupole, where they are filtered according to their mass-to-charge
ratios.
4. The mass-separated ions pass into the T-Wave™ collision cell where
they either undergo collision-induced dissociation (CID) or pass to the
second quadrupole. Any fragment ions are then mass-analyzed by the
second quadrupole.
5. The transmitted ions are detected by the photomultiplier detection
system.
6. The signal is amplified, digitized, and sent to the MassLynx mass
spectrometry software.
Ion optics overview:
Sample cone
Isolation valve
T-Wave collision
cell
Z-Spray ion source
Quadrupole 1
(MS1)
Quadrupole 2
(MS2)
Detector
Conversion dynode
Sample inlet
Transfer optics
Summary of Contents for Xevo TQD
Page 12: ...xii...
Page 20: ...xx Table of Contents...
Page 23: ...Uses and compatibility 1 3 Xevo TQD TP03406...
Page 24: ...1 4 Specifications and Operating Modes Xevo TQD with visor up TP03407...
Page 42: ...1 22 Specifications and Operating Modes...
Page 52: ...2 10 Preparing for Operation...
Page 76: ...3 24 Changing the Mode of Operation...
Page 194: ...4 118 Maintenance Procedures...
Page 208: ...A 14 Safety Advisories...
Page 234: ...B 26 External Connections...
Page 238: ...C 4 Materials of Construction and Compliant Solvents...
Page 250: ...Index 8...