Installing and removing the corona pin
5-13
5. Close the source enclosure.
6. Look through the source window, and use the vernier probe adjuster (see
) to position the ESI probe tip so that it is pointing,
approximately, midway between the tips of the sample cone and corona
pin.
Removing the corona pin from the source
Required materials
Chemical-resistant, powder-free gloves
To remove the corona pin from the source:
1. Prepare the instrument for working on the source (see
2. Pull the source enclosure release (located at the bottom, right-hand side)
outwards, and swing open the enclosure.
3. Remove the corona pin from its mounting contact (see the figure on
).
Tip:
Store the corona pin in a safe location.
4. Fit the blanking plug to the corona pin mounting contact (see the figure
).
5. Close the source enclosure.
Warning:
The LC system connections, ESI probe, and source can
be contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
To avoid electric shock, ensure that the instrument is
prepared for working on the source when commencing this procedure.
Warning:
The source can be hot. To avoid burn injuries, take great
care while working with the instrument’s source enclosure open.
Warning:
The ESI probe tip is sharp. To avoid puncture wounds,
take great care while working with the source enclosure open if an
ESI probe is fitted.
Summary of Contents for Xevo G2 QTof
Page 10: ...x...
Page 18: ...xviii Table of Contents...
Page 34: ...1 16 Waters Xevo G2 QTof Overview...
Page 44: ...2 10 Preparing the Mass Spectrometer for Operation...
Page 227: ...Connecting the liquid waste line B 17 Positioning of drain tube TP01807 Correct Incorrect...
Page 238: ...B 28 External Connections...
Page 242: ...C 4 Materials of Construction and Compatible Solvents...
Page 248: ...Index 6...