Cleaning the extraction cone
5-45
Fitting the ion block assembly to the source assembly
Required materials
•
Chemical-resistant, powder-free gloves
•
Combined 2.5-mm Allen wrench and cone extraction tool
To fit the ion block assembly to the source assembly:
1. Fit the ion block assembly to the PEEK ion block support.
2. Use the combined 2.5-mm Allen wrench and cone extraction tool to fit
and then slowly tighten the 4 ion block assembly securing screws
sequentially and in small increments.
3. Ensure that the source isolation valve is closed (see
4. Fit the source enclosure (see
).
5. Connect the power cords for the mass spectrometer and roughing pump
to the main power source.
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
To avoid puncture wounds, take great care while working with
the source enclosure open if one or both of these conditions apply:
• An ESI probe is fitted (the probe tip is sharp).
• A corona pin is fitted (the pin tip is sharp).
Caution:
To avoid recontaminating the ion block assembly, wear clean
chemical-resistant, powder-free gloves during this procedure.
Summary of Contents for Xevo G2 QTof
Page 10: ...x...
Page 18: ...xviii Table of Contents...
Page 34: ...1 16 Waters Xevo G2 QTof Overview...
Page 44: ...2 10 Preparing the Mass Spectrometer for Operation...
Page 227: ...Connecting the liquid waste line B 17 Positioning of drain tube TP01807 Correct Incorrect...
Page 238: ...B 28 External Connections...
Page 242: ...C 4 Materials of Construction and Compatible Solvents...
Page 248: ...Index 6...