5-10
Maintenance Procedures
Fitting the source enclosure to the instrument
Required materials
Chemical-resistant, powder-free gloves
To fit the source enclosure to the instrument:
1. Using both hands, fit the source enclosure to the two supporting studs
on the source adaptor housing.
2. Close the source enclosure.
3. Connect the probe adjuster and options cables to the instrument’s
connectors.
Tip:
The cables and connectors are color coded; the blue-sleeved cable
connects to the blue connector and the yellow-sleeved cable to the yellow
connector.
4. Slide closed the instrument’s source interface door.
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
To avoid puncture wounds, take great care while fitting the
source enclosure to the source if a corona pin is fitted (the pin tip is
sharp).
Caution:
If you are fitting a NanoLockSpray source enclosure, to
avoid damage to the sample inlet, you must move the sprayer
platform out of the source enclosure before you close the source
enclosure (see
).