Calibrating
63
4
4.2.6 Automatic Calibration Dialog Box
You should perform all three types of calibration: static, scanning, and scan speed
compensation. This lets you subsequently use any data acquisition mode. It also lets you
change mass ranges and scan speeds while maintaining correct mass assignment.
1.
Select
Calibration > Calibrate Instrument
from the Tune window. The Calibration
window appears (
2.
Select
Calibrate > Start Acquisition
from the Calibration window. The Automatic
Calibration dialog box appears (
Figure 4-9 Automatic Calibration Dialog Box
3.
Specify the type or types of calibration you intend to perform:
Static
,
Scanning
,
and/or
Scan Speed
.
Note:
For a complete calibration, select
Static Calibration
,
Scanning Calibration
,
and
Scan Speed Compensation
in the Types area. Then, in the Process area,
select
Acquire & Calibrate
and
Print Report
.
Considerations
Though Waters suggests you perform all three types of calibration, you can nevertheless
specify one, or any combination, of calibration types. Beware, however, that doing so
invokes the following limitations:
•
Specify
Static
to calibrate only for acquisitions where the quadrupole “parks” at a
single mass (for example, SIR acquisitions).
Summary of Contents for Micromass ZQ 2000
Page 6: ......