
3-18
Using the Detector
Preparing to start a run
You must set up a run before you make fluorescence measurements. To start a
run, you can press Run/Stop or trigger the detector’s operation via the inject-
start terminals on the rear panel. When you start a run, the shutter opens
automatically and the detector performs an auto-zero function (when the
function is enabled).
You must select Auto Zero-On-Inject for the detector to automatically auto
While under Empower or Millennium
32
closes when the detector finishes a run, and the run timer stops and then
resets. If you run the detector manually, you can close the shutter by pressing
Reset or waiting until the run ends.
To avoid the automatic opening and closing of the shutter, you can disable the
idle mode by checking the Selection box in the Configuration screen (see
).
Setting up a run
After you press HOME
to return to the home screen and select a channel mode
(
λ
or
λλ
), you can set up the detector for a run. Before beginning a run,
however, you must select the channel mode, and program the following
parameters:
•
Operating wavelength pairs
Tip:
The emission
λ
setting (e
λ
) must be at least 10 nm above the
excitation
λ
setting (x
λ
).
•
Gain
•
Output sensitivity, EUFS
•
Filter type
•
Time constant
•
Analog output type
•
Data units
If you want to perform other functions during a run, you must program other
parameters.
“Accessing primary and secondary functions” on page 3-19
and
table titled “Primary and secondary function (method) parameters” on
Summary of Contents for 2475
Page 14: ...xiv Table of Contents...
Page 158: ...4 16 Maintenance Procedures...
Page 182: ...5 24 Error Messages Diagnostic Tests and Troubleshooting...
Page 196: ...A 14 Safety Advisories...
Page 200: ...B 4 Specifications...
Page 214: ...C 14 Solvent Considerations...
Page 228: ...Index 14...