June 11, 2015, 715004754 Rev. A
Page 22
1.2.3.2
Inhomogeneities in solution
The differential refractometer measures the difference in refraction between the
reference solvent and the sample solution. Most common inhomogeneity problems
are due to improper solvent preparation.
•
The mobile phase must be of constant composition.
•
Gradient separations are not recommended since the
Δ
n
from the mobile phase
change is larger than that from the analyte peak.
Because the 2414 RI detector is sensitive to contaminants, replacing tubing can often
require long wash times. See
1.3 Detector description
The detector is designed for high-performance liquid chromatography applications. It
can operate as a standalone unit with an integrator or with a Waters system controller
or Waters data system.
for system specifications, and to
for solvent
considerations.
Figure 1–5:
2414 RI detector
1.3.1 Range and attenuation
The detector functions with solvents with refractive indices between 1.00 and 1.75.
The measurement range of the instrument is 7
×
10
–9
to 5
×
10
–4
refractive index units
full scale (RIU-FS mode).
1.3.2 Features
The detector has these key features:
•
Standalone programmability – Stores up to 10 user-defined programs (or
methods) consisting of up to 16, programmable, timed events and 1 threshold
event (in standalone RIU mode).
Summary of Contents for 2414
Page 2: ...June 11 2015 715004754 Rev A Page ii...
Page 14: ...June 11 2015 715004754 Rev A Page xiv...
Page 42: ...June 11 2015 715004754 Rev A Page 42...
Page 56: ...June 11 2015 715004754 Rev A Page 56...
Page 62: ...June 11 2015 715004754 Rev A Page 62...
Page 92: ...June 11 2015 715004754 Rev A Page 92...
Page 114: ...June 11 2015 715004754 Rev A Page 114...