Replacing the roughing pump oil
December 6, 2012, 71500126802IVD Rev. C
135
Replacing the air filter behind the source probe
To replace the air filter behind the source probe
1. Remove the probe from the source, see
.
2. Grasp the air filter tab and remove the air filter frame by lifting it
toward you (see the figure on
3. Remove the air filter from the air filter frame and dispose of it.
Removing air filter from frame
4. Insert the new air filter in the air filter frame.
5. Reinstall the air filter.
6. Fit the probe to the source, see
.
Replacing the roughing pump oil
Change the roughing pump oil annually.
Note:
This procedure is not required for an Alcatel oil-free roughing pump.
Required materials:
•
Chemical-resistant, powder-free gloves
•
8-mm Allen wrench
•
Flat-blade screwdriver
TP02640
Air filter
Frame