17
Table 204
Guidance and Manufacturer’s declaration — Electromagnetic Immunity
• Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Model 610 is used
exceeds the applicable RF compliance level above, the Model 610 should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the Model 610.
• Over the frequency range 150 kHz to 80MHz, field strengths should be less than 3V/m.