www.watchgas.eu

D e t e c t i o n

User Manual

 

UNI Sustainable Single Gas Detector

D e t e c t i o n

2

3

Description

 The  UNI Sustainable is a single sensor, portable, personal toxic gas monitor. It displays gas 
concentration continuously on a big segment LCD. It also monitors STEL, TWA, peak and min (for O2 
only) value of the gas and can be displayed on demand. High, Low, STEL&TWA alarm threshold values 
are configurable. The shell is made of high strength, durable material. Two key operation, simple to 
use. Sensor and battery can be replaced easily. Calibration is also very convenient.

 Warning

This manual must be carefully read by all individuals who have or will have the responsibility of using, 
maintaining or servicing this product. The product will perform as designed only if it is used, maintain-
ed and serviced in accordance with the manufacturer’s instructions.

•  Never operate the monitor when the cover is removed.
•  Remove the monitor cover and battery only in area known as non-hazardous.
•  Use only WatchGas lithium battery.
•  This instrument has not been tested in an explosive gas/air atmosphere having an oxygen 

concentration greater than 21%.

•  Substitution of components will impair suitability for intrinsic safety.
•  Substitution of components will void warranty.
•  It is recommended to bump test with a known concentration gas to confirm the instrument is 

functioning properly before use.

•  Before use, ensure that the ESD film on the display is not damaged or peeling.

Proper Disposal

The Waste Electrical and Electronic Equipment(WEEE) directive (2012/19/EU) is intended to promote 
recycling of electrical and electronic equipment and their components at end of life . This symbol 
(crossed-out wheeled bin) indicates separate collection of waste electrical and electronic equipment 
in the EU countries. This product may contain one or more Nickel-metal hydride (NiMH), Lithium-ion, 
or Alkaline batteries. Specific battery information is given in this user guide. Batteries must be recycled 
or disposed of properly. At the end of its life, this product must undergo separate collection and 
recycling from general or household waste. Please use the return and collection system available in 
your country for the disposal of this product.

Summary of Contents for UNI Sustainable

Page 1: ...www watchgas eu UNI Sustainable Single Gas Detector User Manual D e t e c t i o n...

Page 2: ...3 2 2 Sensor Calibration 7 3 2 3 Calibration Adapter 7 3 2 4 Zero Calibration 8 3 2 5 SPAN Calibration 8 3 2 6 Bump Test 8 3 2 7 Change Alarm Limit 8 3 2 8 Change Bump CAL Interval 9 3 2 9 Change SPAN...

Page 3: ...thium battery This instrument has not been tested in an explosive gas air atmosphere having an oxygen concentration greater than 21 Substitution of components will impair suitability for intrinsic saf...

Page 4: ...window 3 LCD 4 Left Key Confirm Number increasing 5 Right Key Power Cursor moving 6 Alligator clip 7 Sensor 8 Vibrator DISPLAY SYMBOLS 1 Gas name includes CO H2S SO2 O2 VOC 2 Question mark 3 OK 4 Gas...

Page 5: ...WARM UP SEQUENCE After powered on the unit enters a warm up and self test sequence and shows the firmware version as follows If the sensor is not able to be identified by the instrument or is not inst...

Page 6: ...from any other screen when there s no key action for 60 seconds 3 1 1 MIN SCREEN FOR OXYGEN SENSOR ONLY 3 1 2 TURN UNIT OFF In normal display mode press and hold the Right Key then the unit will disp...

Page 7: ...use the Right Key to move cursor Once all four digits are input the cursor will move to OK use the Left Key to finish password input and enter the Config mode If the digit input is mistaken use the Ri...

Page 8: ...to the unit then press the Left Key to start span calibration Unit will start count down the count down time depends on different sensors normally 60 seconds after count down is finished span calibra...

Page 9: ...NGE SPAN VALUE The span calibration preset value can also be changed the change process is similar with the alarm limit But the new span will not take effect until user successfully completes a span c...

Page 10: ...status is displayed on the LCD switching between and if the fast startup is enabled or switching between and if the fast startup is disabled Use the Right Key to change the enable disable status use t...

Page 11: ...ws by turning them counterclockwise 4 Remove the top cover after carefully unplugging the buzzer connector 5 Peel a filter from the sheet and center it over the sensor Gently press down 6 Plug in the...

Page 12: ...i o n 12 Filter need to be replaced when it becomes discolored clogged with particles or draws in liquid otherwise it will contaminate sensor and void warranty Caution Change battery only in area kno...

Page 13: ...H ALARM Buzzer 3 beeps per second LED 3 flashes per second Vibration 1 per second HIGH 2 flashes per second LOW ALARM Buzzer 2 beeps per second LED 2 flashes per second Vibration 1 per second LOW 2 fl...

Page 14: ...TY BATTERY ALARM Buzzer 1 beep per minute LED 1 flash per minute Vibration 1 per minute and Battery icon 1 flash per minute 6 Trouble shooting Problem Possible reason Solution Can not turn on unit Bat...

Page 15: ...zero and span Power on zero calibration with user confirmati on Event log Up to 50 alarm events Battery Replaceable AA Lithium battery for 3 years of operation Measurement Diffusion Housing Polycarbo...

Page 16: ...5 W00 NO2 Nitrogen Dioxide 0 20 ppm 38 mg m3 0 1 ppm 0 2 mg m3 M001 0011 W00 H2 Hydrogen 0 1000 ppm 0 2000 ppm 83 mg m3 166 mg m3 1 ppm 1 ppm 0 1 mg m3 0 1 mg m3 M001 0018 W00 M001 0019 W00 PH3 Phosph...

Page 17: ...luding damage by fire lightening water damage or other hazard voids liability of the manufacturer In the event that a product should fail to perform up to manufacturer specifications during the applic...

Reviews: