background image

www.watchgas.nl 

AirWatch 

Quick reference card

D e t e c t i o n

After an alarm, we recommend having the
surroundings inspected by an authorized gas 
measurer/analyst. The AirWatch can only be reset 
after the concentrations measured have dropped 
below the alarm values. The red LED is on 
continuously and the green LED is flashing. 
This reset is achieved by holding down the alarmreset but-
ton.
The green LED will come on and the attached 
machine can be restarted.

Turning the AirWatch off. 
The AirWatch may only be turned off when 
no concentrations of explosive or toxic 
substances are present.
1. Keep the on/off button pushed in for 
approximately two seconds.
2. Enter code 4321.
3. Select > and press the confirmation 
button ‘on/off’.

Alarm 

When a dangerous concentration of gas 
is detected by the AirWatch, it will break 
a potential-free contact which, by means of 
a M.S.A. (Motor Stop Automatic) plug, can 
turn off a diesel-operated or electrical switch 
cabinet. A red LED on the AirWatch indicates 
that there is an alarm.

                 Air Watch      v256
1                   SN: 7-00000

LEL        CO     H2S      OXY
  OK       OK       OK       OK

Calibration   finished!
  OK       OK       OK       OK

Enter  shutdown   PIN
  PIN:   4321   >

Error  calibrating
    0         0       0 .0        20.9

LEL        CO     H2S      OXY
    0         0     0 .0      20.9

             Air Watch          176s
  LEL      CO       H2S      OXY

             Air Watch          176s
            Warming   up ...

Calibrating.. (90s  left)
0         0       0 .0        20.9
Calibrating.. (90s  left)
LEL        CO     H2S      OXY

LEL        CO     H2S      OXY
  OK       OK       OK       OK

LEL        CO     H2S      OXY
CAL        CAL    CAL             CAL

LEL        CO     H2S      OXY
    0         0     0 .0      20.9

FreshAir calibration?
NO                         YES

Enter Unit ID:
            000  >

Setting    accepted
         

      Calibration   menu
>Config        Cal      Exit

Select  option    
UnitID      Pumpcal    Exit

Calibration   menu
  PIN:   1234   >

Calibration   finished!
  OK       OK       OK       OK

Enter menu
PIN:   1234   >

  Select    option
            Config      >Cal      Exit

      AirWatch    cal    menu
>Fresh     Bump    Span    Exit

              Apply  fresh  air
ABORT                        START

Calibrating.. (90s  left)
0         0       0 .0        20.9
Calibrating.. (90s  left)
LEL        CO     H2S      OXY

                   Select        sensors
>LEL       >CO      >H2S        >O2

                   Select        sensors
>LEL       >CO      >H2S        >O2

                   Select        sensors
>LEL       >CO      >H2S        >O2

Bump  result:  OK
    50     50     10.0      18.0

Bump  result:  OK
    please    remove     gas

      AirWatch    cal    menu
    Fresh   >Bump    Span    Exit

      Apply    bump    gas
ABORT                        START

bumping...    ( 30s  left )
50         50     10 .0    18.0

bumping...    ( 30s  left )
LEL        CO     H2S      OXY

Enter menu
PIN:   1234   >

  Select    option
            Config      >Cal      Exit

      AirWatch    cal    menu
    Fresh     Bump   >Span    Exit

      Apply    span    gas
ABORT                        START

Calibrating.. (90s  left)
50         50     10 .0    18.0

Calibrating.. (90s  left)
LEL        CO     H2S      OXY

Enter menu
PIN:   1234   >

  Select    option
            Config      >Cal      Exit

Calibration   finished!
  OK       OK       OK       OK

Calibration   finished!
    please    remove     gas

LEL        CO       H2S      OXY
FAIL      FAIL    FAIL    FAIL

LEL        CO     H2S      OXY
  0         NEG    0.0      21.9

Measuring modus / Meetmodus

Troubleshooting / probleemoplossing.

Change unit ID / Aanpassen unit ID

Fresh air calibration / Verselucht kalibratie

Bumptest

Span calibration modus / Span kalibratie modus

Starting up / Opstarten

LEL        CO     H2S      OXY
  OK       OK       OK       OK
LEL        CO     H2S      OXY
    0         0     0 .0      20.9

Calibrate    pump?
ABORT                        START

  Calibrating   .  .  .
Do not  block  pump         

      Calibration   menu
>Config        Cal      Exit

Select  option    
UnitID      Pumpcal    Exit

Calibration   menu
  PIN:   1234   >

Calibrate pump / Pomp kalibratie

LEL        CO     H2S      OXY
  OK       OK       OK       OK
LEL        CO     H2S      OXY
    0         0     0 .0      20.9

    Verifying   firmware  .  .

���������������

           AirWatch
Upgrade   complete   !

           Update  firmware?
NO         255 ->256        YES

      Upgrade   in progress

�����������

Update Firmware

LEL        CO     H2S      OXY
  OK       OK       OK       OK
LEL        CO     H2S      OXY
    0         0     0 .0      20.9

Calibration   finished

LEL        CO     H2S      OXY
  OK       OK       OK       OK
LEL        CO     H2S      OXY
    0         0     0 .0      20.9

LEL        CO     H2S      OXY
  OK       OK       OK       OK
LEL        CO     H2S      OXY
    0         0     0 .0      20.9

!  !        PUMP  BLOCKED        !  !
      OK       OK       OK       OK

LEL        CO     H2S      OXY
    0         0     0 .0      20.9

Bump  result:  FAIL
      0       0     0.0      20.9

Bump  result:  FAIL
    please    remove     gas

1

2

3

4

5

LEL        CO     H2S      OXY
    0         0     0 .0      LOW

AirWatch           v256

INTERNAL    ERROR

AirWatch           v256

RECONFIGURE   UNIT

CONNECTED      TO      PC

Configuration modus

 AirWatch   programmer   r09

    Verifying   firmware  .  .

���������������

           AirWatch
Upgrade   complete   !

           Update  firmware?
NO         255 ->256        YES

      Upgrade   in progress

�����������

Update Firmware

 AirWatch   programmer   r09

Red LED. The AirWatch alarm is activated. 

A gas concentration is being detected that is  

higher than the alarm values, or the pump is 

blocked.

Red LED is on continuously and green LED 

is flashing. The AirWatch can be reset with the 

alarm/reset button.

Green LED. The AirWatch is in measurement 

mode.

Reviews: