![WATANABE WSPA-F2AL Operating Instructions Download Page 1](http://html1.mh-extra.com/html/watanabe/wspa-f2al/wspa-f2al_operating-instructions_3501830001.webp)
6-16-19, JINGUMAE, SHIBUYA-KU, TOKYO 150-0001, JAPAN
TEL: +81-3-3400-6147 FAX: +81-3-3409-3156
1/4
IM0873
-
00
Operating Instructions
WSPA-F2AL
WSPA-F2ALW
FREE SPEC 2-INPUT ARITHMETIC
AND LOGIC UNIT
We thank you for purchasing the product manufactured by
Watanabe Electric Industry Co., Ltd. Please use the product in
accordance with this manual after confirming that the contents
on the rating label of the product conform to the specifications
you required.
This instruction manual explains how to handle, connect,
and adjust this product.
We are sure that you will be satisfied with this product
because this product has been manufactured and inspected in
accordance with the strict quality control standards. In case of
trouble such as damage during transportation, please contact
your dealer or us as soon as possible.
○
Package Includes
● Arithmetic and logic unit 1 unit
●
Socket 1 piece
1. Product Outline
This unit is a small type plug-in converter designed to
convert two direct current or voltage signals into a direct
current or voltage signal that depends on various arithmetic
processes provided to the inputs including addition and
subtraction.
This product is a digital type CPU-equipped arithmetic and
logic unit. The dedicated setting tool of the unit allows users
to change the primary input range, secondary input range,
output range, arithmetic item, arithmetic expression, and
parameters and perform monitoring as well as provide a
simulated input/output in the field.
* The WSPA-specific setting tool and dedicated USB cable
are separately required to change the settings of this
product. Refer to the instruction manual for the WSPA
setting tool for further information.
2. Functions and Features
●
The primary input range, secondary input range, output
range, arithmetic item, arithmetic expression, and
parameters are changeable at your choice.
●
Up to 5 arithmetic processes available by multi-computing
system.
●
Realization of 5-year guarantee owing to long life design
●
Universal power supply supported
●
CE Marking compliance
●
RoHS Directive compliance
3. Cautions
1) When this product is used as a CE Directive-compliant
item:
●
This product complies with Installation Category II,
Pollution Degree 2, and Maximum Operating Voltage of
300 volts required by CE Directive. Basic insulation is
provided for insulation capability between power supply
terminal and input
・
output of signals. Make sure that the
insulation class of this product meets the required
specification before installation.
● Follow the definition that this product is designed to be
installed in a control panel.
● This product shall be used at an altitude equal to or
below 2000 meters.
● Appropriate space and creeping distance shall be
provided. Inappropriate wiring may revoke compliance
with CE Marking.
CE Directive: Electromagnetic Compatibility Directive
EMI EN61000-6-4
EMS EN61000-6-2
The wiring length should be not more than
30 m
.
Low Voltage Directive EN61010-1
RoHS
Directive EN50581
● Insulation resistance:
Among Input-primary output-secondary output-power
supply terminal
100 M ohms or more/500 V DC
●
Withstand voltage:
Among Input-primary output-secondary output-power
supply terminal
2000 V AC for 1 minute
2) Power supply
●
Check the rating label.
①
For rating: 100-240 V AC
100-240 V AC±10% (50/60 Hz),
approx. 5.5 VA (F2AL), approx. 7 VA (F2ALW)
②
For rating: 24 V DC
24 V DC±10%, approx. 100 mA (F2AL),
approx. 130 mA (F2ALW)
③
For rating: 100-120 V DC
100-120 V DC±10%, approx. 25 mA (F2AL),
approx. 30 mA (F2ALW)
3) When handling
●
When this product is connected to or disconnected from
the socket, be sure to turn off the power and stop the
input signal for safety.
●
When a screw on this product is touched or the
output control switch is operated, be sure to touch
any metal object to release static electricity
beforehand.
4) Installation
●
Be sure to use this product indoors.
●
If this product is installed in a place with a lot of dust or
metal particles, use a dust-proofing case with heat
releasing measures to keep this product clean.
●
Avoid vibration or shock to this product as much as
possible. Failure to observe this may cause trouble.
●
Install this product within the operating temperature
range of –5 to 55 degrees C.
●
Install this product with the operating humidity of 90%
RH or less (no freezing, no condensation).
●
Do not block the ventilation slot of this product.
5) Wiring
● Keep the power supply line, input line, and output line
away from the noise sources, relay drive line, and high
frequency line.
●
Avoid bundling the line of this product together with the
line of superimposed noise and holding the line in the
same duct in which the line of superimposed noise is
held.
6) Setting change with use of the setting tool
●
The setting tool of this product allows users to change
the settings of the parameters, and the settings are stored
in the internal memory.
(The same holds true for the output adjustment value.
Refer to Section 9 for further information.)
The write limit to the internal memory is 10,000 times at
minimum.