background image

17

www.warneronline.com

Using the Model Membrance 

 

Test Procedures

In the following testing procedures, allow a tolerance of ±1% on the readings taken. For 

example, if the test response is indicated as 100 mV, a reading from 99.0 to 101.0 mV 

would be within tolerance. 

Offset Controls

Vm OFFSET (VOLTAGE ELECTRODE section): The full range of this control is ±200 

mV.  This can be verified by rotating the control first fully clockwise and then fully 

counterclockwise while observing the VOLTAGE ELECTRODE METER. The displayed 

readings will indicate off scale at the extremes of the control’s manipulation since the 

meter is only capable of displaying ±199.9 mV. 
Vm x10 output BNC (VOLTAGE ELECTRODE section): This output can be monitored using 

an oscilloscope. The reported voltage will swing between ±2 V as the Vm OFFSET control 

is manipulated throughout its full range. Set the Vm OFFSET to 0.0 reading on the meter 

and verify that the Vm x10 reading on the scope also reads 0 V.
Ve OFFSET control (CURRENT ELECTRODE section): The Ve OFFSET control is tested 

in the same manner as the Vm OFFSET control. This control is adjustable when the 

CLAMP MODE switch is set to off. The voltage of the Ve OFFSET is read from the BATH 

ELECTRODE METER when the CLAMP MODE switch is set to off. 
Ve x10 output BNC: This BNC is located on the rear panel of the instrument and reports 

the setting of the Ve OFFSET control when the CLAMP MODE switch is set to off. 
Return all controls to their initial settings when done with this test.

Voltage Electrode Test

In actual practice, the voltage electrode test is used prior to entering the cell and 

indicates the resistance of the electrode. When used in conjunction with the model cell, 

it measures both the electrode and membrane resistance (2 MΩ). This test is performed 

using controls in the VOLTAGE ELECTRODE section.
With the model cell in place, depress the Vm ELECTRODE TEST pushbutton and observe 

a reading of 20 mV on the VOLTAGE ELECTRODE METER. This corresponds to a 2 MΩ 

reading at a calibrated response of 10 mV/MΩ. 
On the oscilloscope, the Vm x10 output will read 200 mV which is x10 the applied test 

voltage. Since the test current is being passed through 1 MΩ (Rm), the I MONITOR 

output will indicate 1 mV, which corresponds to 10 nA of current.
With the CLAMP MODE switch set to off, the BATH ELECTRODE METER monitors Ve 

(voltage at the current electrode). In this case, Ve will be a measure of the voltage across 

Rm and the meter will indicate 10 mV (1 MΩ). The Ve x10 output BNC on the rear panel 

can also be checked to see that it reads 100 mV (meter reading x10).

 

Summary of Contents for OC-725C

Page 1: ...Publication 5720 001 REV 1 0 Oocyte Clamp Amplifier User s Manual Warner Instruments OC 725C...

Page 2: ...ty Compensation Overload Alarm DC Offsets Nomenclature Text conventions Control Description Front panel Voltage electrode Bath electrodes Clamp section Commands Current electrode Rear panel Additional...

Page 3: ...placement Single oocyte setup with indirect ground Single oocyte setup with direct ground Dual ooctye set up Cable connections Resting position of the controls Using the gain select Other gain range...

Page 4: ...BJECT Comments and Recommendations Membrane damage Repeated recordings Electrophysiology Appendix Specifications Equipment is intended to be operated in a controlled laboratory environment Gain telegr...

Page 5: ...ve limitation may not apply to you If a defect arises within the one year warranty period promptly contact Warner Instruments Inc 1125 Dixwell Ave Hamden Connecticut 06514 5014 using our toll free num...

Page 6: ...grounded Make Proper Connections Make sure all connections are made properly and securely Any signal wire connections to the unit must be no longer than 3 meters Observe All Terminal Ratings Review th...

Page 7: ...aturation The unique design of the bath clamp eliminates the need for series resistance compensation It provides an accurate measurement of bath current by creating a virtual ground in the bath while...

Page 8: ...s with a minimum of leakage Electrode Test for voltage and current electrodes Capacity Compensation for the Vm voltage input Overload Alarm serves as a reminder when the feedback amplifier reaches its...

Page 9: ...conventions which are specified below Since our goal is to provide clarity rather than complexity we welcome any feedback you may wish to provide Warner Instrument product numbers are presented using...

Page 10: ...hbuttons and the Vm output bnc Controls for negative capacity compensation are also located in this block The voltage meter reports the membrane voltage Vm with a full scale range of 199 9 mV The volt...

Page 11: ...he voltage Ve of the current electrode when the clamp mode selector switch is off see clamp section A lit LED indicates voltage readings in mV When in clamp mode clamp mode selector switch set to slow...

Page 12: ...ot required The slow clamp speed is approximately 0 5 ms when measured with the model membrane 1 M shunted with 0 47 F In this mode measured currents are displayed on the BATH ELECTRODES METER in unit...

Page 13: ...N connector for the current electrode Ve offset With a range of 200 mV center zero this control is used to adjust the offset voltage of the current electrode Use this control to establish a zero refer...

Page 14: ...located on the rear panel CHASSIS is common with the instrument enclosure and connected to earth through the power line cord A shorting link allows for interconnection of the two grounds In most exper...

Page 15: ...with two screw mounting slots is attached to the probe base The BATH PROBE connects to the control unit with a 6 pin connector Current Electrode Cable A two meter shielded cable is supplied with a 2 m...

Page 16: ...ent chassis must be directly connected to earth ground High Voltage Outputs When handling the current electrode cable be sure to set the GAIN CONTROL fully clockwise and the CLAMP MODE switch to off C...

Page 17: ...y the 10 k resistors Initial Instrument Settings Connect the model cell to the OC 725C as shown on its cover Be sure to connect the ground wire to the ground mini jack on the side of the bath probe Co...

Page 18: ...is read from the BATH ELECTRODE METER when the CLAMP MODE switch is set to off Ve x10 output BNC This BNC is located on the rear panel of the instrument and reports the setting of the Ve OFFSET contro...

Page 19: ...ge at the current electrode In this case Ve will be a measure of the voltage across Rm and the meter will indicate 10 mV 1 M DC Clamp Test This test is performed using controls in the CLAMP ELECTRODE...

Page 20: ...me of the oscilloscope trace become faster since the speed of the clamp is limited by the resistance of the current electrode and the capacitance of the oocyte If ringing oscillation is observed decre...

Page 21: ...ylgard They will need to be broken off however to a relatively large diameter to insure a fast response time by the clamp For the voltage electrode the pipette tip should be broken back to an O D of 3...

Page 22: ...the mounting rod supplied NOTE Do not fill the current holder with KCl solution Bath Probe The bath clamp is designed to maintain a virtual ground in the oocyte perfusate The bath probe should be posi...

Page 23: ...urrent electrode It is recommended that the user not directly expose the electrode wires to the perfusate if the recording session is to last for more than a few minutes Instead agar bridges should be...

Page 24: ...with the optional DIFFERENTIAL VOLTAGE HEADSTAGE Single Oocyte Setup with Direct Ground Applications where use of the bath clamp is not suitable such as those with a very long solution path to ground...

Page 25: ...lloscope computer or a chart recorder the following connections should be made Vm Membrane potential may be recorded from the Vm x10 connector VOLTAGE ELECTRODE section Ve The voltage of the current e...

Page 26: ...election changes the sensitivity of the current meter Typically measured currents will fall in the x1 range The lower and higher ranges are intended to cover those applications where currents are beyo...

Page 27: ...odified model cell All readings were recorded at 1 kHz 8 pole Bessel and are RMS A lower value feedback resistor increases the speed of the bath clamp and can handle larger currents without saturating...

Page 28: ...f there is a bubble in the pipette which will cause an open circuit 5 Using the Vm OFFSET control adjust the VOLTAGE ELECTRODE potential to read 0 mV If the junction potential of the voltage electrode...

Page 29: ...ance the voltage electrode until it pops through the membrane 12 The potential across the membrane will now be displayed on the VOLTAGE ELECTRODE METER Current Electrode Placement Advance the current...

Page 30: ...GAIN mode CLAMP section may be required to clamp high conductance low resistance cells This condition will be evidenced by the inability of the instrument to maintain a DC holding potential to within...

Page 31: ...sition fully counter clockwise to click off as described above and the CLAMP MODE selector switch be placed in the off position before removing the current electrode from the cell Failure to perform t...

Page 32: ...y setting a dip switch on the main circuit board 1 First disconnect the power cord from the wall 2 Remove the two screws at the rear of the top cover and it off 3 Locate dip switch S10 on the circuit...

Page 33: ...he same resistance as they had in the previous recording A significantly higher resistance could indicate that the pipette is partially plugged with cellular debris Make the following control settings...

Page 34: ...and UV filtered if you plan to use it during recording Minimally the recording chamber can be a stable surface on which the oocyte will not roll around A disposable petri dish with a piece of nylon me...

Page 35: ...F Vm Meter Range full scale 199 9 mV Bath electrode channel Im Ve OFFSET 200 mV Noise Im Clamp clamp on 5 5 nA RMS at 1 kHz x1 range Open loop clamp off 28 pA RMS at 1 kHz x1 range I MONITOR 0 01 100...

Page 36: ...0 1 V command attenuated by 10 Maximum external input 10 V Power requirements 100 120 v 0 5A or 220 240 VAC 50 60 Hz Fuse 0 5A 120v slow blow 0 15A 200v slow blow Appendix Dimensions Enclosure 9 x 42...

Page 37: ...1 4 V 2 1 6 V 5 1 8 V 10 2 0 V 20 2 2 V 50 2 4 V 100 2 6 V Gain Select Headstage Resistor Im Output Range V A Im max output A Maximum Meter Reading x0 1 10 k 0 01 1 0 10 1000 199 9 A x1 0 100 k 0 1 10...

Page 38: ...Xenopus oocytes Transcription and Translation eds B D Hames and S J Higgins IRL Press Oxford Ch 10 Hille B 1984 Ionic Channels of Excitable Membranes Sinauer Sunderland MA Ch 2 Zhou J Potts J F Trimme...

Page 39: ...38 Publication 5720 001 REV1 0 Notes...

Page 40: ...39 www warneronline com Notes...

Page 41: ...40 Publication 5720 001 REV1 0 Notes...

Page 42: ......

Page 43: ...746 Phone 508 893 8999 Toll Free 800 272 2775 Fax 508 429 5732 E mail support hbiosci com Web www harvardapparatus com Germany Hugo Sachs Elektronik Gruenstrasse 1 March Hugstetten D 79232 Germany Pho...

Reviews: