Rev 01.19.04
Set temp = 35 °C
Flow rate (ml/min)
Monitored temp (°C )
Output voltage (V)
1 34.0
4.44
2 33.4
5.20
3 32.5
5.72
4 31.7
6.10
5 31.0
6.86
Set temp = 37 °C
Flow rate (ml/min)
Monitored temp (°C )
Output voltages (V)
1 35.9
4.85
2 35.6
5.53
3 34.6
6.31
4 34.1
7.03
5 33.2
7.46
Set temp = 40 °C
Flow Rate (ml/min)
Monitored Temp (°C )
Output voltages (V)
1 38.9
5.22
2 38.2
6.13
3 37.3
6.87
4 36.3
7.45
5 36.7
7.52
Set temp = 45 °C
Flow Rate (ml/min)
Monitored Temp (°C )
Output voltages (V)
1 43.2
6.15
2 42.4
6.88
3 41.9
7.72
4 41.4
8.37
5 40.9
8.75
Set temp = 50 °C
Flow Rate (ml/min)
Monitored Temp (°C )
Output voltages (V)
1 47.9
6.89
2 47.4
7.85
3 47.2
8.95
4 46.2
9.69
5 44.6
10.3
Warner Instruments, Inc.
1125 Dixwell Avenue, Hamden, CT 06514 USA
•
(203) 776-0664
•
(800) 599-4203
•
Fax (203) 776-1278