background image

Rev 01.19.04

 

 

 

Set temp = 35 °C 

Flow rate (ml/min) 

Monitored temp (°C ) 

Output voltage (V) 

1 34.0 

4.44 

2 33.4 

5.20 

3 32.5 

5.72 

4 31.7 

6.10 

5 31.0 

6.86 

 

 

 

Set temp = 37 °C 

Flow rate (ml/min) 

Monitored temp (°C ) 

Output voltages (V) 

1 35.9 

4.85 

2 35.6 

5.53 

3 34.6 

6.31 

4 34.1 

7.03 

5 33.2 

7.46 

 

 

 

Set temp = 40 °C 

Flow Rate (ml/min) 

Monitored Temp (°C ) 

Output voltages (V) 

1 38.9 

5.22 

2 38.2 

6.13 

3 37.3 

6.87 

4 36.3 

7.45 

5 36.7 

7.52 

 

 

 

Set temp = 45 °C 

Flow Rate (ml/min) 

Monitored Temp (°C ) 

Output voltages (V) 

1 43.2 

6.15 

2 42.4 

6.88 

3 41.9 

7.72 

4 41.4 

8.37 

5 40.9 

8.75 

 

 

 

Set temp = 50 °C 

Flow Rate (ml/min) 

Monitored Temp (°C ) 

Output voltages (V) 

1 47.9 

6.89 

2 47.4 

7.85 

3 47.2 

8.95 

4 46.2 

9.69 

5 44.6 

10.3 

 

 