6
E.
INJECTION PRESSURE:
Displays injection pressure in either PSI or kPa (maximum 60 PSI; 413
kPa).
F.
PRESSURE UNIT toggle switch:
Toggle switch controls injection pressure display unit of
measure. Injection pressure can be displayed in either pound-force per square inch (PSI) or
kilopascals (kPa).
G.
PRESSURE OUTPUT (P
out
) connector:
Pressurized gas output line. Output tubing line
connects P
out
to injection pipette holder.
H.
MANUAL INJECTION TRIGGER:
Press-button trigger delivers regulated pressure for the
digitally set period of time at the desired pressure.
I.
EXTERNAL INJECTION TRIGGER INPUT BNC:
This input is for electrically initiating injection
from an external trigger source. A positive TTL pulse of is required (baseline 0 V, trigger 5 V,
lasting >10 ms).
J.
INJECTION TIME SETTING DIAL:
This dial controls the injection time setting. Turning the dial
changes the injection time in seconds; pressing the dial in and turning changes the injection time
in fractions of a second.
K.
INJECTION PRESSURE REGULATOR:
This seven turn control sets the injection pressure over
the range from 0.2 to 60 PSI (1.4 to 413 kPa).
L.
POWER switch:
This switch turns the instrument on and off.
Rear panel
Warner Instruments
•
1125 Dixwell Avenue, Hamden, CT 06514 USA
(203) 776-0664
•
(800) 599-4203
•
Fax (203) 776-1278
M.
PRESSURE INPUT (P ) connector:
This connector is the input for the compressed gas. 75 PSI
in
is the recommended input pressure; maximum input pressure is 105 PSI.
Rev 090714