Evidence of the medical device’s basic suitability for effective sterilization was provided by an independent
test laboratory using the LISA 517 B17L* steam sterilizer (W&H Sterilization S.r.l., Brusaporto (BG)) and the
CertoClav MultiControl MC2-S09S273** steam sterilizer (CertoClav GmbH, Traun).
> “Dynamic-air-removal prevacuum cycle” (type B): 134°C (273°F) – 3 minutes*, 132°C (270°F) –
4 minutes*/**
> “Gravity-displacement cycle” (type N): 121°C (250°F) – 30 minutes**
Drying times:
> “Dynamic-air-removal prevacuum cycle” (type B): 132°C (270°F) – 30 minutes**
> “Gravity-displacement cycle” (type N): 121°C (250°F) – 30 minutes**
* EN 13060, EN 285, ISO 17665
** ANSI/AAMI ST55, ANSI/AAMI ST79