Maintenance
LISA | Instructions for Use | VA131 2019 ENG Rev06 | 16/04/2020 | © 2020 W&H Sterilization Srl
81
CLEANING THE CHAMBER AND THE CHAMBER ACCESSORIES
Proceed as follows:
1
Remove the trays and the chamber rack.
2
Clean the chamber with a damp sponge and a mild detergent solution paying
attention not to bend or damage the temperature probe inside the sterilizer
chamber.
3
Rinse with water.
4
Clean the trays and the chamber rack with a damp sponge and a mild detergent
solution.
5
Rinse with water.
6
Reposition all pieces of the chamber accessories properly.
Note: The trays and the tray holder may also be cleaned in a washer disinfector.