Technical data
Endea
EB-75
EB-79
Transmission ratio
16:1
2:1
Coupling
ISO 3964
Recommended root canal instruments*
Instrument shaft diameter (Ttyp 1)
according to standard ISO 1797
(mm)
NiTi-files for rotation root canal preparation
Ø 2.35
min. chuck length
(mm)
engaging
max. motor speed for NiTi files from 300-350 rpm:
produces an application speed of: (rpm)
5,000
312
600
300
max. motor speed for NiTi files from 600 rpm:
produces an application speed of: (rpm)
10,000
625
1,200
600
max. motor speed for NiTi files from 1,200-2,500 rpm:
produces an application speed of: (rpm)
25,000
1,562
2,500
1,250
5,000
2,500
*
In choosing the correct operating conditions, the user must ensure that there is no risk to the user, the patient or third parties.
Always follow the instructions of the manufacturer of the root canal instruments (e.g. in terms of speed, chuck length,
described application).
37