background image

 
2 INSTALLATION   

 

 

 

 

 

9070 

 

 
 

9

 

With the transmitter case secured to a wall or panel, open the 
chassis to the case and wire the main power connection.  Fit the 
connection blocks onto the circuit boards. 

9

 

Fit the chassis unit, secure it with the captive screws and replace 
any ground (earth) bonding leads.  

9

 

Position the face panel and secure it with the four plastic fasteners. 

 

2.2 SAMPLE 

REQUIREMENTS 

 

 

[

Warning:

  Do not exceed the maximum pressures and temperatures 

specified. If you use pressure-reducing equipment, install a pressure relief 
valve between your equipment and the sample inlet to the monitor for safety 
purposes. 

Bring the sample to the temperature and pressure suitable for measurement, 
using sample coolers and pressure-reducing equipment (see Section 8)

 

2.3 EXTERNAL 

PIPING 

CONNECTIONS 

2.3.1

 INLET

 

 

The sample should be connected to the sensor unit using ¼” o.d. tubing of 
stainless steel or rigid plastic.  Connect this to the sample inlet coupling on the 
right hand side of the case’s bottom piece. 

 

The inlet tube should be of sufficient wall thickness to withstand the highest 
sample pressure, and pipe lengths should be kept short.  

 

Where particulate matter is present (e.g. magnetite in boiler samples), fit a 60-
micron sample filter to the sample line.  

 

A shut-off valve (not supplied with the monitor) is necessary in the sample 
unit. 

2.3.2 DRAIN 

 

The drain from the cup at the bottom of the sensor unit case consists of a hose 
barb suitable for 3/8” bore plastic or rubber tubing.  Alkaline 

 

effluent from the monitor and overflowing sample drains through this single 
connection. 

Summary of Contents for AQUALERT mAI-9070 Series

Page 1: ...Waltron LLC AQUALERT DIVISION ___________________________________________ Water Chemistry Measurement Control AI 9070 SERIES HYDRAZINE MONITOR INSTRUCTION MANUAL...

Page 2: ...s shown in Fig 4 1 14 4 2 QA QC JUNCTION BOX SWITCH 15 4 2 1 Normal Calibration 15 4 2 2 Process Calibration 15 4 2 3 QA QC 15 4 3 TRANSMITTER UNIT 16 4 3 1 ELECTRONICS CHASSIS 16 4 3 2 CIRCUIT BOARD...

Page 3: ...y flashes reading out 30 6 4 3 Display shows CF 30 6 4 4 Display flashes reading continuously 30 6 4 5 Display shows Hot 30 6 4 6 CALIBRATION FAIL ALARM 30 6 4 7 MALFUNCTIONS OF THE MONITOR 31 REPLACE...

Page 4: ...ows for either calibration on a process sample or a quality analysis quality control QA QC test The liquid handling section contains a machined PVC head chamber As sample flows to the vapor entrainmen...

Page 5: ...t light emitting diodes l e d which show the level of hydrazine and operational information as well The operator can see when the instrument is in the calibration mode and when the calibration has not...

Page 6: ...radiant heat sunlight and drafts Avoid areas containing chlorinating equipment Mount the sensor unit no more than 33 feet 10 meters from its associated sample coolers Mount the sensor unit on the wal...

Page 7: ...NSTALLATION 2 1 2 TRANSMITTER UNITS Wall mount using four 5 16 or bolts on 9 06 by 9 06 centers Sufficient access space as detailed previously must be left for making cable connections to the transmit...

Page 8: ...eners turn in either direction 9 Remove the chassis 9 Secure the transmitter case to the wall by four 5 16 or bolts Note Before fitting the chassis unit and connecting it to the main power supply chec...

Page 9: ...t using sample coolers and pressure reducing equipment see Section 8 2 3 EXTERNAL PIPING CONNECTIONS 2 3 1 INLET The sample should be connected to the sensor unit using o d tubing of stainless steel o...

Page 10: ...opening the junction box which could admit moisture You may mount the transmitter either next to the sensor unit or up to 330 feet 100 meters away Also allow sufficient space on the transmitter unit...

Page 11: ...shown in Fig 2 4 You may pull the terminal block away from the pins on the board if required FIGURE 2 4 WIRING THE INTERCONNECT CABLE TO TRANSMITTER Caution The terminal block for the sensor unit con...

Page 12: ...2 INSTALLATION 9070 Analog Board TB1 Analog Board TB2 Analog Board TB3 9...

Page 13: ...The choice of two different isolated recorder output signals enables you to use the instrument with a wide variety of recording and data processing equipment The load requirements are shown in Sectio...

Page 14: ...resistors as shown in the two examples in Fig 2 5 Our example consists of four watt resistors A suitable resistor network kit is listed in Section 7 Your external equipment should be set up using the...

Page 15: ...ge tip into the gel filling hole Make sure the air vent hole is unobstructed Depress the plunger of the syringe to force gel into the round space between the sensor and outer jackets Continue pressing...

Page 16: ...r is now in operation The lamp adjacent to the number display indicates the hydrazine level units of measurement in ppb or the sample temperature in degrees Celsius 9 If required turn the range switch...

Page 17: ...he constant head unit Here the head unit stabilizes the effect of changes in sample pressure and flow rate When the monitor loses sample a small tube overflowing into the constant head on one side ens...

Page 18: ...the pressure relief valve will open The valve will allow the sample to pass through the heat exchanger thus bringing the standard solution to a similar value to the sample solution 4 2 QA QC JUNCTION...

Page 19: ...the display 9 Analog Board Middle board containing the analog input and current output circuitry 9 PSU Board Rear board containing the power supply and output relays 4 3 2 CIRCUIT BOARD FUNCTION The m...

Page 20: ...ntrols mounted on the front circuit board in the chassis protrude through holes in the faceplate Four plastic fasteners secure the chassis to the faceplate A three digit red l e d display shows the hy...

Page 21: ...razine level being measured within the overall instrument range Remote range indication is provided 2 SET VALUE These buttons and arrows increase or decrease the value displayed on the digital indicat...

Page 22: ...2 Use in conjunction with the SET VALUE and arrows buttons to set the values at which the alarm relays operate 6 TEMP Pressing STD and CAL together gives an indication of the sample temperature in C 7...

Page 23: ...osed If the power source fails both external alarms sound to indicate a malfunction 9 Two other sets of relay contacts are provided One set changes over during a calibration sequence and the other set...

Page 24: ...ly 9 In AUTO as the concentration increases switching to the next range takes place at 100 of the current output giving a 50 output on the upper range 10 in linear output When the concentration decrea...

Page 25: ...ainers with high purity water and fill them with fresh standard solution Perform a calibration weekly In cases where a weekly calibration is not necessary determine a suitable schedule to suit the ope...

Page 26: ...5 CALIBRATION PROCEDURE 9070 5 2 CALIBRATION USING PROCESS SAMPLE AS A STANDARD SOLUTION 9 Place junction box selector switch to PROC CAL position 23...

Page 27: ...as it causes burns and irritation to eyes respiratory system and skin Wear rubber gloves and eye protection In warm weather pressure increases in the bulk container of ammonia Release the cap careful...

Page 28: ...e dilution The following dilution method produces more accurate results than those obtained by calorimetric laboratory analysis of the final diluted standard 9 Pipette 10ml of the 1000ppm stock soluti...

Page 29: ...volume Store in a plastic container Warning Prepare the nitric acid solution under a fume hood and take the appropriate precautions when handling concentrated acids 6 1 4 HYDRAZINE SENSOR RECHARGE GEL...

Page 30: ...l drain 9 Carefully pull the sensor apart and wash the components thoroughly to remove all traces of gel 9 Withdraw the platinum electrode taking care not to damage the electrode or the electrical con...

Page 31: ...n into the socket at the top of the liquid handling panel 9 Open the shut off valve upstream of the sensor unit and adjust until sample is overflowing from the constant head unit 9 Carry out a calibra...

Page 32: ...ve the sensor from the sample panel 9 Disassemble sensor outer jacket from center ceramic assembly 9 Clean the sensor of any gel 9 The sensor can be stored as a dry item 6 4 UNSCHEDULED SERVICING Abno...

Page 33: ...M Calibration Fail condition occurs after a CALIBRATION as the electronics fail to see sufficient output from the hydrazine sensor A number of factors requiring investigation could cause this event 9...

Page 34: ...stment of the high purity sample Systematically check mechanical components involved with liquid handling for leaks or blockages these problems change the chemical conditions around the sensor Most pr...

Page 35: ...6 MAINTENANCE 9070 FIGURE 6 2 PART NUMBERS OF PLASTIC TUBING 32...

Page 36: ...try 8 point Sample Sequencer K1148 004 4 point Manifolded Sampling Valve K1148 008 8 point Manifolded Sampling Valve K1138 633 8 point Sample Inlet Connections 1 8 OD Swagelok K1138 634 8 point Sample...

Page 37: ...nal standard 20ppb minimum Calibration Time 5 minutes Ambient Temp 32 to 131 F 0 to 55 C Enclosure Finish Corrosion resistant polyurethane Battery Backup 10 years 8 1 Installation Information Sample T...

Page 38: ...l Conn via access glands as required Power Supply 100 120 200 240Vac Requirements 50 60 Hz 100VA Power Supply Voltage 10 20 Tolerances Frequency min 47Hz max 65Hz Case Protection Transmitter Unit IP55...

Page 39: ...9 OPERATOR S NOTES 9070 9 OPERATOR S NOTES 36...

Page 40: ...nltd com Waltron s technical expertise and extensive experience has enabled us to provide personalized solutions to the water quality industry It is Waltron s commitment to provide you with timely and...

Page 41: ...followed when operating the analyzers in conditions of high pressure and or temperature 9 The chemicals that are used to operate this machine must be stored away from heat protected from temperature...

Page 42: ...artment within the company returning the items for repair 9 The name telephone number and extension of the individual in the company responsible for returning items for repair 9 A brief description of...

Page 43: ...t liability through alleged failure or failures of any of its products or their accessories International All international orders will have a warranty period of one year after installation or 18 mont...

Page 44: ...kages making sure that the number is in agreement with the shipping paperwork If there are any outward signs of damage please retain all containers and packages for inspection by carrier It is general...

Reviews: