Maintenance
45
Waltron User Manual 101-016-B.3
3041 Silica Analyzer
Annually
3 ways valve cleaning
Colorimetric cell o-ring replacement
Fittings cleaning / replacement
Mixing loop tubing replacement
Reagent straw and sample tubing replacement
Analyzer general inspection (for qualified personnel only)
Waltron provides all of the tubing and fittings required for scheduled maintenance as part of the
yearly consumables kit, part number W3040-101.
Table 7.1: Scheduled manintenance reference chart.
Scheduled Maintenance
Frequency
Daily
Monthly
Quarterly
Annually
Visual check of FAULT alarm indicator
X
Visual check of liquids enclosures for leakages
detection
X
Visual check of reagents level
X
Sample fast-loop reservoir cleaning
X
Colorimetric cell cleaning
X
Reagents replacement (reset reagent counters)
X
Pinch valve tubing replacement
X
Sample pump tubing replacement
X
Reagent pumps tubing replacement
X
3 way valve cleaning and fittings replacement
X
Mixing loop tubing replacement
X
Colorimetric cell O-ring replacement
X
Reagent straw and sample tubing replacement
X
Analyzer general inspection (for qualified
personnel only)
X