background image

101-016-B.3 

 

 

EXPERTS IN WATER CHEMISTRY SINCE 1903

 

 

 

 

 

 
 

3041 Silica Analyzer 

User Manual

 

Summary of Contents for 3041

Page 1: ...101 016 B 3 EXPERTS IN WATER CHEMISTRY SINCE 1903 3041 Silica Analyzer User Manual...

Page 2: ...5546 www waltron net Please be ready to provide the following information Date analyzer was purchased Analyzer model and serial number Recent maintenance history Calibration slope values and detailed...

Page 3: ...iners packages and chemicals Only qualified personnel should be involved in the installation operation and servicing of the analyzer Follow safety precautions when operating analyzer in conditions of...

Page 4: ...al Service Department Customer name address and department Name and telephone number of the individual responsible for returning items for repair Brief problem description Ship to Waltron service cent...

Page 5: ...g and shipping needs Check all items received against those on the packing list Chemicals are usually shipped in a separate package and will be itemized accordingly Verify that the number of packages...

Page 6: ...1 2 1 General Information 10 1 2 2 List of Warnings and Potential Dangers 11 1 2 3 Reagents 12 1 2 4 Sample 12 1 2 5 Waste Disposal 12 1 2 6 Analyzer General Hazards 12 2 Introduction 15 2 1 Analyzer...

Page 7: ...2 Chart 35 5 6 3 Manual Step 35 5 7 Data Logger 36 5 7 1 Save to USB 36 5 8 Program Menu 37 5 8 1 Settings Menu 37 5 8 2 Calibration Menu 38 5 9 Service Menu 39 6 Operation 40 6 1 Reagents Prime 40 6...

Page 8: ...3 2 Reccomended complete system installation 26 Figure 3 3 Electrical connections for user connections and power connections 28 Figure 3 4 Mounting schematics for the reagent holding bracket 29 Figur...

Page 9: ...etermination 18 Table 3 1 Terminal block pin locations 28 Table 5 1 User password 31 Table 5 2 Menu access level with password 31 Table 5 3 Analyzer operating statuses 32 Table 5 4 Main menu button de...

Page 10: ...or 4 configurable alarm relays Reproducibility 1 ppb or 2 of reading whichever is greater Degree of Protection IP55 NEMA 4 Wet section IP65 NEMA 4x Electronics Required Maintenance Monthly replenish r...

Page 11: ...ritten on this manual The user has the responsibility to read and understand the information contained in this manual and to abide by all applicable employee health and safety regulations Use maintena...

Page 12: ...ltic pump motor input terminal mixing pump heater Hazard of chemical burns This symbol is used to indicate the potential hazard of severe burns and serious injury due to hazardous chemicals All handli...

Page 13: ...erials of construction listed below Table 1 2 List of materials used in Waltron 3041 Analyzer Pump tubing Silicon or Norprene Fittings PP Connection tubing Norprene Silicon Colorimetric cell Glass Mix...

Page 14: ...been designed built and located in a closed enclosure with a special key Inside the enclosure these parts have protective covers to avoid any contact and physical injuries to users HAZARD Hazard of b...

Page 15: ...ctrocution have been designed built and located in an enclosure with a special key Inside the enclosure these parts have protective covers with warning labels to avoid contact and serious injuries or...

Page 16: ...colorimetric method The analyzer is assembled in an enclosure with a lockable door The LIQUIDS enclosure includes all the components involved in sample and reagent flows as well as mixing and reactio...

Page 17: ...is measured at a specific wavelength The absorbance is related to sample concentration according to Beer s law The Lambert Beer law is an empirical relationship that relates the absorption of light t...

Page 18: ...ges from 0 to 1 but it can go higher An absorbance of 0 at some wavelength means that no light of that particular wavelength has been absorbed The intensities of the sample and reference beam are both...

Page 19: ...wait functions Following the addition of Reagent 2 the mixing pump is activated and the liquid is pumped from the lower part to the upper part of the colorimetric cell The waiting time is programmed...

Page 20: ...4 2 FLOW DIAGRAM Figure 2 4 Flow diagram 2 4 3 ANALYZER OPERATIONS The Waltron 3041 Silica Analyzer performs several operations throughout a cycle A list of common operations is shown below Rinse 1 Ri...

Page 21: ...mple circulation to avoid accumulation of suspended solids The sample flow should be adjusted to allow continuous sample overflow to waste through the U shaped stainless steel tube Inside the reservoi...

Page 22: ...Introduction 21 Waltron User Manual 101 016 B 3 3041 Silica Analyzer Figure 2 6 Mounting dimensions for the fast loop reservoir 10 36 259 mm 5 84 146 mm 5 24 131 mm 5 20 130 mm 0 26 6 5 mm...

Page 23: ...ample inlet s and calibration standard inlet pass through the housing enclosure on the right hand side of the unit and are connected to the 3 way valve Colorimetric Reaction Cell The colorimetric reac...

Page 24: ...s a constant level of few cm of liquid on the colorimetric cell 2 7 ELECTRONICS The microprocessor and its PCB assembly are located in the electronic section It provides control for the entire analyzi...

Page 25: ...tive humidity If the temperature could fall below 5 C the analyzer should be installed in a heated cabinet Because of chemicals and waste gases it is necessary to have a well ventilated location for t...

Page 26: ...t 8 20 cm on either side and about 40 100 cm in the front Sufficient space for the reagent containers should be provided on the side or beneath the analyzer The reagent containers should be positioned...

Page 27: ...26 Waltron User Manual 101 016 B 3 3041 Silica Analyzer Installation Figure 3 2 Reccomended complete system installation Note Single channel analyzers will only have one Fast Loop Reservoir...

Page 28: ...fficiency of the analyzer s grounding Users and qualified maintenance personnel must proceed as follows Be aware of electrical shock and or electrocution labels placed on the analyzer Always isolate p...

Page 29: ...labels starting with pin 1 on the left Figure 3 3 Electrical connections for user connections and power connections Table 3 1 Terminal block pin locations Pin Number Terminal Label Description 1 Inpu...

Page 30: ...below to mount the reagent bracket below the analyzer NOTE Mount the bracket below the analyzer so that the base of the bracket is at least 14 inches 40 cm below the analyzer but no more than 30 inch...

Page 31: ...gle channel analyzers connect the sample to connection 1 on the inlet block and the standard to connection 2 b For dual channel instruments connect sample 1 to connection 1 sample 2 to connection 2 an...

Page 32: ...ressing the touchscreen buttons Figure 5 1 User interface main display screen 5 2 USER PASSWORD Table 5 1 User password First Level Password service 1111 Second Level Password admin Contact Waltron Ta...

Page 33: ...ROGRAM menu Analysis Cycle and started when the START ON LINE button has been pressed As soon as the analyzer completes an analysis cycle it will restart with a new analysis This condition is identifi...

Page 34: ...vel falls below a value of 4 the analyzer goes into FAULT status and stops automatically The alarm message will be displayed in the main menu and the FAULT contact will be activated Every time the rea...

Page 35: ...de and wait for the user s next command CH 2 available on dual stream units only This command will perform a single analysis cycle on stream 2 At the end of the cycle the analyzer will go into stand b...

Page 36: ...ted Blank shows the most recent blank value Led shows the led supplied Valid it shows the validation value in compared to the calibration value Current operation this window displays the current analy...

Page 37: ...results in memory Figure 5 6 Data logger screen of the user interface 5 7 1 SAVE TO USB To export the stored results in the data log to a csv file 1 Insert a USB memory stick into the bottom of the t...

Page 38: ...to set the ratio of the analysis cycles versus the extra cycles the alarm value low or high and to enable or disable the loss of sample alarm and the 4 20 mA output CALIBRATION MENU This function is...

Page 39: ...een analysis cycles when the analyzer is in on line mode Result alarm Programs the low high value alarm The alarm is expressed in of full scale Output Signal Enables disables the 4 20 mA output Datalo...

Page 40: ...ange of the analyzer This setting only applies to dual stream analyzers Configuring Relay 1 2 3 and 4 There are 4 cells on the service screen to configure Relay 1 Relay 2 Relay 3 and Relay 4 Touch one...

Page 41: ...ion for 20 seconds 4 Select ADD REAG 1 2 3 functions for 60 seconds each in order to prime the three reagent lines Visually confirm that reagent is being drawn into the lines and into the flowcell If...

Page 42: ...od of extended shut down The recommended and default setting for automatic calibration is approximately once per week 1 Login with the 1st level service password 1111 2 Connect the CAL tube to the cal...

Page 43: ...w to the fast loop reservoir 2 Check that the sample fills the fast loop reservoir so that the float switch rises and sample is being discharged from the overflow drain 3 Put the analyzer into online...

Page 44: ...r user command 7 When the user is finished with the grab sample analysis make sure to reconnect the sample tubiing to the fast loop resevoir 8 Turn on sample flow to the fast loop resevoir 9 Put the a...

Page 45: ...enance work all necessary precautions regarding personal safety protective clothing safety glasses etc are to be taken into consideration Always be sure to label and rinse all connected tubing with wa...

Page 46: ...nintenance reference chart Scheduled Maintenance Frequency Daily Monthly Quarterly Annually Visual check of FAULT alarm indicator X Visual check of liquids enclosures for leakages detection X Visual c...

Page 47: ...s 7 Place the pump half containing the rotor in one hand and place the rollers in the 2 6 and 10 o clock positions Place tubing in the outer port and against the two rollers as shown keeping your thum...

Page 48: ...drain tubing into the drain valve pinch mechanism 9 Feed the drain tubing through the bottom of the enclosure 10 Connect the drain tubing to the T fitting 11 Perform a manual command of sample 1 for...

Page 49: ...first pump roller into the open space 9 From the manual commands menu perform an add reagent 1 command for 1 second This should rotate the first pump roller into place against the tubing and open a s...

Page 50: ...eriod please proceed as follows 1 Replace the reagents in the container with distilled water and perform a priming of all the pumps with DI water a Login with the 1st level service password 1111 b Pre...

Page 51: ...cell Loss of Sample 1 The sample to stream 1 is lost Check the flow of Sample 1 Loss of Sample 2 The sample to stream 2 is lost Check the flow of Sample 2 Dual stream analyzers only Emergency Stop So...

Page 52: ...are Parts Listing Description Part Number Pump Head 7016 Sampling Pump W9040 007 Pump motor 70 80 rpm 24 Vdc W3040 015 External Reservoir Fast loop with level switch W3040 003 Peristaltic pump for rea...

Page 53: ...52 Waltron User Manual 101 016 B 3 3041 Silica Analyzer Spare Parts Display Cable W3040 093 Output Cable for User Connections W3040 094...

Page 54: ...Appendix A Panel Mount Schematics 53 Waltron User Manual 101 016 B 3 3041 Silica Analyzer APPENDIX A PANEL MOUNT SCHEMATICS...

Page 55: ...54 Waltron User Manual 101 016 B 3 3041 Silica Analyzer Appendix A Panel Mount Schematics...

Reviews: