5.3

 

Temperature Measurement 

5.3.1

 

The  instrument  can  be  used  to  measure  temperature 
independently with the temperature probe. 

5.3.2

 

When  measuring  temperature  independently,  remove 
the pH electrode connected to the instrument. 

5.3.3

 

Press  the  READ/AUTO  key  for  3  seconds  to  activate 
continuous  reading  mode.    Error  code  may  appear  on 
pH display but will not affect the temperature reading. 

5.3.4

 

Allow some time for the temperature reading to stabilize 
before taking the measurement value. 

 

5.4

 

mV Measurement 

The  instrument  mV  measuring  function  allows  ORP,  ion-
selective mV and others mV measurements. 

5.4.1

 

Replace  the  pH  electrode  with  the  measuring 
electrode.  

5.4.2

 

Rinse  the  measuring  electrode  in  distilled  water  and 
switch on the instrument. 

5.4.3

 

Rinse  the  measuring  electrode  in  test  sample  solution 
and  then  in  another  test  sample  solution  and  press  the 
READ/AUTO key. 

5.4.4

 

The  Wait  indicator  will  flash  momentarily  until  reading  is 
stabilized.  The instrument will beep and the ‘AUTOLOCK’ 
indicator appears.  Reading is established. 

5.4.5

 

To disable the AUTOLOCK feature and allow continuous 
reading,  press  and  hold  the  READ/AUTO  key  for  3 
seconds  until  a  beep  sound  is  heard.  To  resume  to 
AUTOLOCK mode, press the READ/AUTO key once. 

5.4.6

 

If reading is erratic or unstable, the instrument will not be 
able to AUTOLOCK the reading.  This can be the result of 
a worn-off or bad electrode. 

11 

Summary of Contents for HP9000

Page 1: ...Operations Manual WalkLAB pH mV ORP temperature meter HP9000 16 1 16 1 2 15 2 15 14 3 14 3 4 13 4 13 12 5 12 5 6 11 6 11 10 7 10 7 8 9 8 9...

Page 2: ...ARTED 6 3 1 Unpacking 6 3 2 Setting Up 6 4 CALIBRATION PROCEDURE 6 4 1 Calibration with Temperature Probe 7 4 2 Calibration without Temperature Probe 8 5 MEASUREMENT 9 5 1 pH Measurement with Temperat...

Page 3: ...it requires very low power in OFF and AUTOLOCK mode Ease of use Automatically lock reading when a stable reading is established Feature can be disabled to allow continuous reading display Automatic sh...

Page 4: ...tomatic Temperature 0 to 99 9 C Compensation Buffer Recognition pH 7 pH 4 or pH 10 Buffer Temperature Range 0 to 60 C Electrode Slope Recognition 20 at pH 4 or pH 10 25 C Electrode Offset 60 mV at pH...

Page 5: ...2 FRONT PANEL DESCRIPTION 3 Main Display Temperature Display Temperature Probe pH Electrode...

Page 6: ...ON OFF key turns the instrument on and off When switch ON display will show the present reading in AUTOLOCK mode MODE The MODE key selects reading from pH to mV display During calibration MODE select...

Page 7: ...r press and hold the key 2 3 ERROR CODES ER 1 pH out of range pH less than 0 or more than 14 00 pH mV out of range mV less than 999 or above 999 mV ER 2 Calibration buffer solution s temperature below...

Page 8: ...pH electrode into their respective connectors on the side of the instrument 3 2 3 Next you must calibrate the instrument Follow the calibration procedure in step 4 Always switch on the instrument for...

Page 9: ...performing a multiple data sampling It will take a little longer time for an accurate calibration point 4 1 5 When OFFSET point is established the instrument will beep and the SLOPE indicator will fl...

Page 10: ...00 rinsing solution and then into the pH 7 00 calibration solution and press the CAL key The OFFSET indicator will flash momentarily Once the instrument detected the correct offset buffer the OFFSET...

Page 11: ...le solutions one for rinsing and the other for testing This will prevent carry over of solution during testing 5 1 pH Measurement with Temperature Probe 5 1 1 Rinse the pH electrode and temperature pr...

Page 12: ...next digit will flash Select the digit you wish to change while it flashes and press DIGIT ADV key to advance the digit Digit will advance at each press of the key Digit will not advance by double pre...

Page 13: ...ments 5 4 1 Replace the pH electrode with the measuring electrode 5 4 2 Rinse the measuring electrode in distilled water and switch on the instrument 5 4 3 Rinse the measuring electrode in test sample...

Page 14: ...11 4 00 10 32 5 7 08 4 00 10 25 10 7 06 4 00 10 18 15 7 03 4 00 10 12 20 7 01 4 00 10 06 25 7 00 4 01 10 01 30 6 98 4 02 9 97 35 6 98 4 02 9 93 40 6 97 4 03 9 89 45 6 97 4 04 9 86 50 6 97 4 06 9 83 55...

Page 15: ...within warranty will be repaired at no charge Make sure that the instrument is properly packed and insured against possible damage or loss in shipment Purchase invoice MUST accompany returned product...

Page 16: ...and work procedures and the result of inspection or testing for calibration meets the product specifications above The standard solutions use for the calibration procedure was tested by instrument wh...

Reviews: