40
ELECTROMAGNETIC COMPATIBILITY (EMC)
• Guidance and manufacturer’s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should only be used in
such environments:
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
RF energy is used only to maintain device’s operation. Therefore,
its RF emissions are so low that it’s not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The device is suitable for use in all establishments, including
domestic establishments, and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Complies
• Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be used in
such environments:
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment–
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact discharge
± 15 kV air discharge
± 8 kV contact discharge
± 15 kV air discharge
In air discharge testing, the climatic
conditions shall be within the
following rages:
Temperature:15˚~35˚,
Humidity:30%~60%.
Power frequency
(50 or 60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
50 or 60 Hz
3 A/m
50 or 60Hz
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 2 kV Power lines
± 2 kV Power lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0% UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270°, and 315°
0% UT; 1 cycles
70% UT; 25/30 cycles
0% UT; 250/300 cycles
0% UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270°, and 315°
0% UT; 1 cycles
70% UT; 25/30 cycles
0% UT; 250/300 cycles
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the device
requires continued operation during
power mains interruptions, it is
recommended that the device be
powered from an uninterruptible power
supply or a battery.