3.3 Connecting the Main unit and Components

Fig. 5

Touch probe

Measuring wire

Main unit

File clip

Lip hook

Handpiece 
base

Motor Handpiece

Single-head 
measuring wire

Contra-angle Handpiece

Fig. 5  Explanation of connecting the main unit

 

3.4 Charge the battery 

When the power indicator flashes, stop using the device and recharge the battery. 
a) Connect the AC adapter with the right side socket of the device properly, and insert the AC 
adapter plug to the power socket. 
b) The green indicator light on the right side is on when charging, and the green indicator light is 
off after charging It may take 4 hours for charging.
Attention: Don't use the device when it is charging.

3.5 Automatic shutdown

When not used for 4 minutes, this device will automatically shut down and all the display and
function will stop.

4. Equipment instructions

4.1 Instructions for root canal length-measurement mode

4.1.1 Length-measurement mode user and Connecting  measuring wire

 

1) Press the ON/OFF button, turn it on.
2) Click the

Apex

on the screen.

3)

, click the sound button icon on the right side of the 

After entering the root canal length mode

screen to turn on / off (    /    ) the tip position prompt sound. This operation is only effective in the 
root canal length Measuring mode

4) Insert the plug of measuring wire into the left side socket of main unit.
5) Insert the file clip and the lip hook into the two sockets of the measuring line

.(

Fig. 6

6) Make the lip hook touch the file clip (as shown in Figure 7), the display 
shows "

 " and reaches the "

" area, indicating that the machine is in normal operation 

 

- -

OVER

(Fig. 7). If there is no display,it means the file clip or measuring wire is damaged and must be 
replaced.

fig.6

fig.7

fig.8

4.1.2 Instruction of the “APEX” Screen Display

A) When the File is far from the apical foramen: the blue indication bars are displayed. (Fig. 9)
b) When the File is near by the apical foramen, the green bars displayed.(Fig. 10)
c) When the file extrudes the apical foramen, the red bars are displayed.(Fig. 11)
d) The number of bars does not represent the distance till the apex, but only the proximity to 
the apex. 

fig

.

9

fig

.

10

fig

.

11

7

8

Summary of Contents for ENDO PRO SMART-TOUCH

Page 1: ...ENDO PRO SMART TOUCH USER MANUAL W A L D E N T TM...

Page 2: ...ting the Main unit and Handpiec 3 2 Connecting the Main unit and Handpiece 6 3 3 Connecting the Main unit and Components 7 3 4 Automatic shutdown 7 4 Equipment instructions 7 4 1 Instructions for root...

Page 3: ...e and pressure to avoid cross infection 1 2 Model Dimensions and Weight of the main unit Model EndoPro Smart Touch 1 3 Structure EndoPro Smart Touch is composed of main unit motor handpiece contra ang...

Page 4: ...1 Product appearance structure ON OFF Button Screen Charging Indicator Charging Port Handpiece port Measuring Port Battery cover 2 The figures of the main accessories Fig 2 i Adapter Fig 2 The figure...

Page 5: ...oper operation of the device 2 5 This device has electromagnetic interference which is similar to other devices patients or doctors with pacemakers are forbidden to use this device 2 6 While being in...

Page 6: ...Press the ON OFF button turn it on 2 Click the Apex on the screen 3 click the sound button icon on the right side of the After entering the root canal length mode screen to turn on off the tip positio...

Page 7: ...s less than 2mm Corner of mouth Lip hook File clip push File handle Metallic part of file Working and moving part of file Do not clamp on this part fig 12 a b c 9 10 When the file reaches the apical p...

Page 8: ...y sign 3 Screen of Recipro and other functions Fig 16 Fig 16 2 Screen of Rotary Fig 15 File Brand Files name Color of files Speed Working mode Change File sizes Torque 11 12 PREMIUM TAPER W A L D E N...

Page 9: ...150rpm to 1000rpm B Torque setting Torque is from 0 6N cm to 4 6N cm Notice If use the function of combination ENDO enlargement with apex locator the contra angle must be covered by the rubber case 4...

Page 10: ...ot canal in a bad Situation 2 Electromagnetic interference 1 Removal of liquid and residual pulp 2 Please check the equipment used nearby and be away from interference Endo file does not reverse rotat...

Page 11: ...ber Attention consult accompanying document Class II equipment Type B applied part Direct current Ordinary equipment 10 Statement All rights of modifying the product are reserved with the manufacturer...

Reviews: