Waikato SYP User Manual Download Page 4

 

 

P4 

Operation 

SYP features a fully configurable sampling regime including two means of collection and 3 

sampling modes (explained below). All modes support a fully configurable purge / rinse cycle 
between samples. 
 

Set up and use is described in following sections. 

Sample method: Gravity & peristaltic pump 

Gravity mode

 was designed for collecting fluid drips falling from a height. It requires a funnel 

or reservoir be set up to collect the liquid, which flows into the machine by gravity alone 

(without any mechanical support). Gravity mode uses full vial detection to ensure consistent 
sample volume and log when samples are completed. 

 
Gravity mode supports continuous, timed or delayed start modes (explained in detail below). 

 

 

 

Pump mode

 was designed for sampling surface or ground water. Here a peristaltic pump 

moves fluid from a reservoir into vials at user-defined intervals. The pump runs for a pre-
determined time for each sample to ensure consistent volume, which is safe-checked by the 

full vial detection feature. 
 

Pump mode supports timed or delayed start modes but not continuous mode. 
 

 

Sample mode: Continuous, timed, delayed start 

Timed

 is the standard mode of operation. A schedule is set by the user dicating sample 

interval from 1 to 65,000 hours. By default the system will purge before a sample is taken. 
 

Delayed start mode

 uses either a user-defined timer or sensor data (air pressure, humidity, 

temperature) to trigger sample collection. Once the user-defined threshold is met, the device 
continues sampling according to user-defined settings. 

 

Continuous

 mode provides an option for a flow-based sampling regime. Water is fed into the 

system and directly into vials. When the system detects a vial is full, it automatically 
timestamps sample completion and immediately begins the next sample. By analysing time 

stamps at each sample, flow rate can be inferred. 
 

Continuous mode only works while using gravity as the means of collection. 

 

 

 

When setting up the device in gravity mode, ensure there is sufficient hydraulic head 

to drive flow to vials by gravity alone. Kinks or dips in tubing will cause operating 
issues or missed samples.

 

 

 

Pump mode can set up with a funnel or reservoir in tight or cramped settings where 
flow to vials by gravity alone is unreliable.

 

 

Summary of Contents for SYP

Page 1: ...SYP fluid sampler User manual...

Page 2: ...c pump 4 Sample mode Continuous timed delayed start 4 Connect to SYP 5 Wi Fi 5 Android OS Phones 5 Apple iOS Phones 5 Windows PC 5 Load the SYP interface 6 6 Configuration and use 7 Status page 7 How...

Page 3: ...ing quality stainless fasteners custom 3D printed components and a transparent acrylic case with carry handles for convenience SYP is the perfect companion for scientists at universities research inst...

Page 4: ...ts timed or delayed start modes but not continuous mode Sample mode Continuous timed delayed start Timed is the standard mode of operation A schedule is set by the user dicating sample interval from 1...

Page 5: ...Use Wi Fi 4 Tap the SYP network in the list Apple iOS Phones 1 Go to Settings 2 Click on Wi Fi Make Sure it is On 3 Choose the SYP to connect Windows PC 1 Make sure your PC has Wi Fi available 2 Click...

Page 6: ...ur phone s camera This works on newer versions of Android and iOS Figure 2 2 Open your web browser Google Chrome is recommended and type in 192 168 4 1 into your search bar While connected to SYP you...

Page 7: ...ng the sampler ensure the settings are set correctly on the Settings page If you want to start at a specific sample enter the sample number in the Start Sample field before pressing Start To start SYP...

Page 8: ...edle block will move to the purge location regardless of where it currently is Note While stopped the Next Sample Time won t be updated Resume This will move the Auto Sampler back into the sampling st...

Page 9: ...P9 Software updates ADD...

Reviews: