21
Wagner & Guder Medical GmbH
issue May 2021
Direction for use, cubeLITE
Manufacturer's Declaration
Electromagnetic Emissions (Tab. 201 according to DIN EN 60601-1-2)
The refraction place cubeLITE is intended for use in an electromagnetic environment
as described below. The customer or user of the refraction place cubeLITE should
ensure that the device is used in such an environment.
Warning! The use of longer cable length may cause an increased emission or a reduced
interference resistance. The use of other sensors or cables except the ones mentioned
above is not allowed.
Cable length incomming voltage ( 2,5m )
Wall mounted box incomming cable ( 3m )
Guidance and manufacturer’s declaration – Electromagnetic emissions
This product is intended for use in the electromagnetic environment specified below. The
customer or the user of this product should assure that it is used in such an environment
Emission test
Compliance
Electromagnetic enviroment -
guidance
RF Emissions
according to CISPR 11
Group 1
This product uses RF energy
only for its internal function.
Therefore, its RF emissions are
very low and are not likely to
cause any interference in
nearby electronic equipment.
RF Emissions
according to CISPR 11
Class B
This product is suitable for use
in all establishments, including
domestic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Emission of harmonics
according to IEC 61000-3-2
Class B
Voltage fluctuations / flicker
emissions according to EN
61000-3-3
Matches