VXI VT1509A User Manual Download Page 7

·

The channel parameter must specify a single channel.

To query the filter state of channel 2 send

INP:FILT?  (@102)

query channel 2

enter statement here 

Querying the

Channel Gain

While the VT1509A’s amplifiers have fixed gain, the channel gain  can be
queried. The response to this query will always be 64. To query any channel 
to determine its gain setting use the INPut:GAIN? (@<channel>) command. 
The INP:GAIN? command returns the current gain value for the specified
channel.

·

The channel parameter must specify a single channel.

To query the gain setting of channel 8 send

INP:GAIN?  (@108)

query channel 8

enter statement here 

Detecting Open

Transducers

This SCP provides a method to detect open transducers. When Open
Transducer Detect (OTD) is enabled, the SCP injects a small current into
the HIGH and LOW input of each channel. The polarity of the current pulls
the HIGH inputs 17 volts and the LOW inputs towards -17 volts. If 
a transducer is open, measuring that channel will return an over-voltage
condition. OTD is available on a per SCP basic. all eight channels of an
SCP are enabled or disabled together. See Figure 5 for a simplified
schematic diagram of the OTD circuit.

Programming With SCPI Commands

VT1509A Fixed Amplifier SCP  7

Signal Conditioning Plug-on

Multiplexer

Signal Input

High

Low

100M

-17V

3K

Gnd.

3K

100M

+17V

High

Low

Figure  5  Open Transducer Detect Circuit

Summary of Contents for VT1509A

Page 1: ...HP E1413s as Agilent HP E1413 Option 19 Enclosed is the User s Manual for the VT1509A Signal Conditioning Plug on Insert this manual in your VT1413C and Agilent HP E1313 manual behind the Signal Cond...

Page 2: ......

Page 3: ...d commands and explains the capabilities of this SCP Finally it covers specifications for this SCP The contents of this manual are Installation 3 Identifying the Plug on 3 Connecting To The Terminal M...

Page 4: ...e stickers are as follows For VT1413C and above Terminal Modules use stickers for VT1509A SCPs The connections are shown in Figure 1 For Agilent HP E1313 Terminal Moduless use stickers for VT1509A SCP...

Page 5: ...dule VT1509A Fixed Filter Amplifier SCP 5 Figure 2 VT1509A B size Terminal Module Connections Ch 00 31 Figure 3 VT1509A B size Terminal Module Connections Ch 32 63 Figure 4 Agilent HP E1413B Option 19...

Page 6: ...s 0 through 7 send SYST CTYP 100 query SCP type ch 0 enter statement here Querying the Filter Cutoff Frequency While the the VT1509A does not provide programmable cutoff frequency the filter frequency...

Page 7: ...8 enter statement here Detecting Open Transducers This SCP provides a method to detect open transducers When Open Transducer Detect OTD is enabled the SCP injects a small current into the HIGH and LO...

Page 8: ...n the SCP The first channel on each SCP is 0 8 16 24 32 40 48 and 56 To enable Open Transducer Detection on all channels on SCPs 1 and 3 DIAG OTD ON 100 116 0 is on SCP 1 and 16 is on SCP3 To disable...

Page 9: ...will be returned to the Query Response Register Detecting Open Transducers Open Transducer Detection OTD is controlled by bits in the Card Control Register For more information on OTD see Figure 1 Ca...

Page 10: ...ired 200 to 850 C Resistance Opt 15 required 128W to 131 kW FS Strain 25 000 e or limit of linear range of strain gage Maximum input voltage Normal mode plus common mode Operating 16 V peak Damage lev...

Page 11: ...cifications The temperature accuracy specifications include instrument and firmware linearization errors The linearization algorithm used is based on the IPTS 68 78 standard transducer curves Add your...

Page 12: ...5 to 65 C 65 to 85 C OFF ON 0 0061 C 0 0055 C 0 011 C 0 010 C 100 W Reference RTD A D Filter 125 to 70 C OFF ON 0 080 C 0 080 C 100 W RTD A D Filter 200 to 75 C 75 to 300 C 300 to 600 C 600 to 970 C O...

Reviews: