5-2
Chest Freezer
VWR International
Section 5
Alarms
This condition is created when the high stage compressor and fans run for
30 minutes and are not capable of cooling the interstage heat exchanger to
the proper temperature. Under this condition, the high stage compressor
and fans will turn off after 30 minutes and an audible and visual alarm will
occur along with the "HS SYST FAIL" message in the LED message
center.
When multiple alarm conditions occur, active messages are displayed in the
message center one at a time, updating at 5 second intervals. Pressing Mute
during multiple alarms causes all active alarms to be silenced and to ring
back in 15 minutes.
An internal communication failure has occurred with the micro board.
During this alarm, the compressor(s) attempt to run continuously.
However, with this type of failure, freezer operation becomes
undependable.
Communication between the micro board and the display board has been
lost. Under this condition, the visual alarm flashes along with dashes in the
temperature display (----). In addition, ‘LOST COMM’ flashes in the
message center. Contact Technical Services.
High Stage System Failure
Alarm
Multiple Alarms
Micro Board Failure
Alarm
Lost Communication
Summary of Contents for VWRC0386A
Page 55: ...8 1 VWR Collection VWR International Section 8 Parts ...
Page 56: ...VWR Collection 8 2 VWR International Section 8 Parts ...
Page 59: ...8 5 VWR Collection VWR International Section 8 Parts ...
Page 60: ...VWR Collection 8 6 VWR International Section 7 Specifications ...
Page 61: ...VWR Collection 8 7 VWR International Section 8 Parts ...
Page 68: ...9 1 VWR Collection VWR International Section 9 Refrigeration Schematics ...
Page 69: ...VWR Collection 9 2 VWR International Section 9 Refirgeration Schematics ...
Page 70: ...VWR Collection 9 3 VWR International Section 9 Refirgeration Schematics ...
Page 71: ...9 4 VWR Collection VWR International Section 9 Refrigeration Schematics ...
Page 72: ...10 1 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 73: ...VWR Collection 10 2 VWR International Section 10 Electrical Schematics ...
Page 74: ...10 3 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 75: ...VWR Collection 10 4 VWR International Section 10 Electrical Schematics ...
Page 76: ...VWR Collection 10 5 VWR International Section 9 Section title ...
Page 77: ...10 6 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 78: ...VWR Collection 10 7 VWR International Section 10 Electrical Schematics ...
Page 79: ...10 8 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 80: ...VWR Collection 10 9 VWR International Section 10 Electrical Schematics ...
Page 81: ...10 10 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 82: ...VWR Collection 10 11 VWR International Section 10 Electrical Schematics ...
Page 83: ...10 12 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 93: ...VWR International D 2 Chest Freezer ...
Page 94: ......