D.O. meter 

OX 4100 H

OPERATING MANUAL

European Catalogue Number:  

OX 4100 H

:662-0176

4100 H

OX

Version: 

2

Issued:

September 11, 2014 9:05 am

Summary of Contents for OX 4100 H

Page 1: ...ba77071e02 09 2014 D O meter OX 4100 H OPERATING MANUAL European Catalogue Number OX 4100 H 662 0176 4100 H OX Version 2 Issued September 11 2014 9 05 am...

Page 2: ...OX 4100 H 2 ba77071e02 09 2014 Legal address of manufacturer VWR International bvba Researchpark Haasrode 2020 Geldenaaksebaan 464 B 3001 Leuven 32 16 385011 http be vwr com Manufactured in Germany...

Page 3: ...Setting the date and time 15 5 Operation 16 5 1 Switching on the meter 16 5 2 General operating principles 17 5 2 1 Operating modes 17 5 2 2 Navigation 18 5 2 3 Navigation example 1 Setting the langu...

Page 4: ...ta transmission 42 5 6 2 Connecting a PC 43 5 7 MultiLab Importer 43 5 8 Reset 44 5 8 1 Resetting the measurement settings 44 5 8 2 Resetting the system settings 45 6 Maintenance cleaning disposal acc...

Page 5: ...ience the operators will know the necessary safety precautions to take when handling chemicals Safety instructions Safety instructions in this operating manual are indicated by the warn ing symbol tri...

Page 6: ...left the factory in a safe and secure technical condition Function and operational safety The smooth functioning and operational safety of the meter can only be guaranteed if the generally applicable...

Page 7: ...ged no longer operates as described in this manual If you are in any doubt please contact the supplier of the meter Obligations of the purchaser The purchaser of this meter must ensure that the follow...

Page 8: ...reliably The OX 4100 H provides the maximum degree of operating comfort reliability and measuring certainty for all applications The proven OxiCal calibration procedure supports you in your work with...

Page 9: ...rement settings F1_ Menu Opens the menu for system settings On Off Switches the meter on or off MODE Selects the measured parameter CAL CAL_ Calls up the calibration procedure Displays the calibration...

Page 10: ...Sensor symbol calibration evaluation calibration interval 5 Measured temperature with unit 6 Status line 7 Softkeys and date time 3 2 4 5 6 7 1 HOLD 25 0 7 9 C O2 mg l 15 03 2014 08 00 Menu USB outpu...

Page 11: ...connect D O sensors to the meter that cannot return any voltages or currents that are not allowed SELV and current circuit with current limiting Almost all customary D O sensors fulfill these conditi...

Page 12: ...ies 4 x 1 5 V alkali manganese batteries type AA Rechargeable batter ies 4 x 1 2 V NiMH rechargeable batteries type AA no charging function Operational life Up to 800 h without 100 h with illumination...

Page 13: ...O partial pressure mbar 0 200 0 0 200 0 1250 0 1 1 1 T C 0 50 0 0 1 Accuracy 1 digit Variable Accuracy Concentration mg l 0 5 of measured value at ambient temperature 5 C 30 C Saturation 0 5 of measur...

Page 14: ...ons CD ROM with USB drivers detailed operating manual Software MultiLab Importer 4 2 Initial commissioning Perform the following activities Insert the supplied batteries Switch on the meter Set the da...

Page 15: ...he meter has an energy saving feature to avoid unnecessary battery depletion The energy saving feature switches off the meter if no key is pressed during the adjusted interval How to set the switch of...

Page 16: ...n 5 3 1 The automatic switchoff function switches off the meter if no key is pressed for an adjustable period The automatic switchoff function is not active if the meter is supplied with power via a c...

Page 17: ...2 1 and section 5 2 2 5 2 1 Operating modes The instrument has the following operating modes Measuring The measurement data of the connected sensor are shown in the measured value display Calibration...

Page 18: ...r settings and dialogs in procedures contain further sub elements The selection is made with the keys The current selection is displayed with a frame Submenus The name of the submenu is displayed at t...

Page 19: ...annot be selected Example Note The principles of navigation are explained in the two following sections by reference of examples Setting the language section 5 2 3 Setting the date and time see sectio...

Page 20: ...The instrument is in the measuring mode 2 Using F1_ Menu open the Storage config menu The instrument is in the setting mode 3 Select the System submenu with The current selection is displayed with a f...

Page 21: ...he setting mode for the Language with OK 8 Select the required language with General Interface Clock Service information Reset System 15 03 2014 08 00 Back Language Deutsch Beep Off Illumination On Co...

Page 22: ...red values Therefore check the time at regular intervals Note After a fall of the supply voltage empty batteries the date and time are reset Setting the date time and date format The date format can b...

Page 23: ...e and confirm the setting with and OK The time is set 7 If necessary set the Date and Date format The setting is made similarly to that of the time 8 To make further settings switch to the next higher...

Page 24: ...pendent system settings to the default con dition Settings To open the Storage config menu press the F1_ Menu key in the measured value display After completing the settings switch to the measured val...

Page 25: ...lity control function see section 5 4 3 The measured parameter flashes on the display as soon as the measured value is outside the stability range when the automatic Stability control is switched off...

Page 26: ...ia as incorrect values would result The USB interface is not galvanically isolated Temperature measurement The D O sensor Oxy 11 has an integrated temperature sensor Preparatory activities Perform the...

Page 27: ...e SAL indicator is displayed in the measuring window Note You can switch the salinity correction on or off and enter the salinity in the menu for calibration and measurement settings see section 5 4 3...

Page 28: ...rol Settings The settings are available in the menu for measurement and calibration settings To open the settings display the required parameter in the measured value display and press the F1 Menu or...

Page 29: ...ith the aid of a comparison measurement e g Winkler titration For details see section 5 4 4 Sal correction On Off Manual salt content correc tion for concentration mea surements Salinity 0 0 70 0 Sali...

Page 30: ...ison measurement by a correc tion factor When the correction multiplier is active the Factor indicator appears in the measuring window Stability control AutoRead In calibration the Stability control f...

Page 31: ...tion vessel must be moist not wet Leave the sensor in the air calibration vessel for a time long enough to adjust Display Calibration record Relative slope S 0 8 1 25 S 0 7 0 8 S 0 6 0 7 Error Error S...

Page 32: ...parame ter flashes 5 Wait for the end of the AutoRead measurement or accept the calibration value with OK The calibration record is displayed and output to the interface 6 Using F1 Continue or OK swit...

Page 33: ...lay press the CAL_ key The calibration records of the last calibration procedures are available in the F1 Menu Calibration Calibration data storage and F1_ Menu Storage config Data storage Calibration...

Page 34: ...tion records with Output the displayed calibration record to the interface with F2 USB output Quit the display with F1 Back or OK Switch directly to the measured value display with MODE Calibration Ca...

Page 35: ...asured value of the connected sensor Measured temperature value of the connected sensor AutoRead info AR appears with the measured value if the AutoRead criterion was met while storing stable measured...

Page 36: ...rocess and switch to the measured value dis play confirm No Then you can e g store the data from the memory to a PC see section 5 5 3 and subsequently erase the memory see section 5 5 4 1 Press the ST...

Page 37: ...the STR_ key The menu for automatic data storing appears Specified entire storing duration Max available storing dura tion Graphical display of the memory usage ID number 1 Interval 30 s Duration 180...

Page 38: ...mode If the automatic storing function is active the meter provides an energy saving mode Eco mode to avoid unnecessary energy consumption The energy saving mode switches off functions of the meter t...

Page 39: ...ory can be shown on the display and output to the interface Editing the data memory The memory is edited in the menu Storage config Data storage To open the Storage config menu press the F1_ Menu key...

Page 40: ...tput the displayed dataset to the interface with F2 USB output Quit the display with F1 Back Data storage Manual data storage Erase Erases the entire manual measurement data memory Note All calibratio...

Page 41: ...splay and move to the next higher menu level with F1 Back 5 5 4 Erasing the measurement data memory How to erase the measurement data memory is described in section 5 5 3 DISPLAYING AND EDITING THE ME...

Page 42: ...on records Data Control Operation description Currentmeasured values of all con nected D O sen sors Manual With F2 USB output Simultaneously with every manual storing process see section 5 5 1 Automat...

Page 43: ...and CD ROM drive Microsoft Windows for details see enclosed installation CD Driver directory 5 7 MultiLab Importer With the aid of the MultiLab Importer software you can record and eval uate measureme...

Page 44: ...ng a reset The following settings for D O measurements are reset to the default settings with the Reset function The sensor settings are reset under the Reset menu item in the menu for calibration and...

Page 45: ...t to the delivery status The system settings are reset in the menu Storage config System Reset To open the Storage config menu press the F1_ Menu key in the measured value display Setting Default sett...

Page 46: ...cing the batteries Caution Make sure that the poles of the batteries are positioned correctly The signs on the batteries must correspond to the signs in the battery compartment Note Alternatively you...

Page 47: ...ust not be disposed of with unsorted waste Instead it s your responsibility to correctly dispose of your equipment at lifecycle end by handling it over to an authorized facility for separate collectio...

Page 48: ...2014 6 5 Accessories Measuring cells Order no pHenomenal OXY 11 OXYGEN SENSOR PHENOMENAL 3M 8 PIN 664 0042 Accessories Order no SURVIVAL KIT PHENOMENAL 662 1166 Extension stick 662 1224 COMMUNICATION...

Page 49: ...nsor contaminated Clean D O sensor and replace it if necessary Cause Remedy Calibration interval expired Recalibrate the measuring system Cause Remedy Batteries almost empty Replace the batteries see...

Page 50: ...What to do if OX 4100 H 50 ba77071e02 09 2014...

Page 51: ...version see page 49 1 Install the downloaded firmware update on a PC An update folder is created in the Windows start menu If an update folder already exists for the meter or meter type the new data i...

Page 52: ...Firmware update OX 4100 H 52 ba77071e02 09 2014...

Page 53: ...measuring system e g the displayed value to the correct value or a value that is regarded as correct Often this expression is also used when the measuring system is adjusted at the same time see adjus...

Page 54: ...etermined by measuring the electrical conductivity Salt content General designation for the quantity of salt dissolved in water Slope relative Relation of the slope value to the value of a theoretical...

Page 55: ...38 F Firmware update 51 I Initial commissioning 14 15 Initialize 44 K Keys 9 M Measured value display 18 Measurement data memory Edit 39 Erase 39 Memory locations 35 Measurement dataset 35 Measuring...

Page 56: ...Lists OX 4100 H 56 ba77071e02 09 2014...

Page 57: ...roduct has been damaged by accident abuse misuse or misapplication or from ordinary wear and tear If the required main tenance and inspection services are not performed according to the manuals and an...

Page 58: ...Compliance with local laws and regulations OX 4100 H 58 ba77071e02 09 2014...

Page 59: ......