6 – VWR STandaRd Line PiPeTTORS 

instruction Manual – 7

 

ambient operating temperature is +5°C to 45°C. 

 

ambient storage conditions (in the original packaging 

during transport and short storage) is -25°C to 55°C.

When working with the pipettor:

 

Follow general work safety regulations regarding  

hazards related to work in the laboratory.

 

Take special care when pipetting aggressive 

substances.

 

Use appropriate protective attire (e.g., clothing,  

goggles, and gloves).

 

avoid pointing the pipettor at yourself or others  

during use.

 

Only use parts and accessories recommended  

by the manufacturer.

 5.  Specifications

The pipettor is a high quality instrument which offers 

excellent accuracy and precision. The accuracy and 

precision (repeatability) of the liquid volume depends on 

the quality of pipet tips used. The values for accuracy 

and precision shown in the table below were obtained 

using manufacturer non-filter pipet tips. Those tips are 

recommended for use to ensure compatibility, accuracy 

and precision when pipetting.

Nominal 

Volume 

(μL) 

Volume 

(μL)

Accuracy 

(%)

Precision 

(%)

Non-filter 

Tips (μL)

Single-channel Pipettors

2

Min.

0.1

±40.0

≤12.0

10

0.2

±12.0

≤6.0

1

±2.7

≤1.3

Max.

2

±1.5

≤0.7

10

Min.

0.5

±4.0

≤2.8

10

1.0

±2.5

≤1.8 

5.0

±1.0

≤0.6

Max.

10.0

±0.5

≤0.4

20

Min.

2

±3.0

≤1.5

200

10

±1.0

≤0.5

Max.

20

±0.8

≤0.3

Nominal 

Volume 

(μL) 

Volume 

(μL)

Accuracy 

(%)

Precision 

(%)

Non-filter 

Tips (μL)

100

Min.

10

±1.6

≤0.80

200

50

±0.8

≤0.24

Max.

100

±0.8

≤0.20

200

Min.

20

±1.2

≤0.60

200

100

±0.8

≤0.25

Max.

200

±0.6

≤0.20

1,000

Min.

100

±1.6

≤0.40

1,000

500

±0.7

≤0.20

Max. 1,000

±0.6

≤0.15

5,000

Min.

500

±1.2

≤0.50

5,000

1,000

±0.6

≤0.25

2,500

±0.6

≤0.20

Max. 5,000

±0.5

≤0.15

10,000

Min. 1,000

±2.5

≤0.6

10,000

5,000

±0.8

≤0.3

Max. 10,000

±0.5

≤0.2

8-channel and 12-channel Pipettors

8 - 10 

12 - 10

Min.

0.5

±10.0

≤8.0

10

1

±8.0

≤6.0

5

±4.0

≤2.0

Max.

10

±2.0

≤1.2

8 - 50 

12 - 50

Min.

5

±4.0

≤2.5

200

25

±3.0

≤1.2

Max.

50

±1.6

≤0.6

8 - 200 

12 - 200

Min.

20

±3.0

≤1.5

200

100

±1.5

≤0.8

Max.

200

±1.0

≤0.6

8 - 300 

12 - 300

Min.

50

±1.6

≤1.5

300

150

±1.2

≤1.0

Max.

300

±1.0

≤0.6

The accuracy and precision were obtained gravi-

metrically using manufacturer tips performing at least 

10 measurements of distilled water at a temperature of 

20°C ± 1°C according to en iSO 8655 standards.
The use of tips from other manufacturers or filter tips 

may result in incorrect liquid aspiration and require 

pipettor recalibration.

Summary of Contents for avantor

Page 1: ...tandard Line Instruction Manual Pipettors VWR Standard Line VWR 8 channel Standard Line VWR 12 channel Standard Line This manual is available in additional languages at www vwr com PMS 647C CMYK 96 54...

Page 2: ...s are designed for microplate filling The pipet tors enable precise and simultaneous delivery of 8 or 12 preset volumes of liquid These pipettors are available in four volume ranges 0 5 10 L 5 50 L 20...

Page 3: ...ti channel manifold Contains a set of plungers and a set of flexible suspension shafts which func tion to reduce the force required to attach the tip Model Identification The volume range of the pipet...

Page 4: ...Accuracy Precision Non filter Tips L Single channel Pipettors 2 Min 0 1 40 0 12 0 10 0 2 12 0 6 0 1 2 7 1 3 Max 2 1 5 0 7 10 Min 0 5 4 0 2 8 10 1 0 2 5 1 8 5 0 1 0 0 6 Max 10 0 0 5 0 4 20 Min 2 3 0 1...

Page 5: ...o not attach the tips with a rocking movement as this may damage the shaft or plunger Observe this rule particularly with single channel pipettors of low volume range Never draw liquids directly into...

Page 6: ...In 2 and 10 L pipettors place the cap that can be purchased separately on the bottom of the pipettor shaft and slide the cap upwards to the bottom of the tip ejector M L 10000 L L 10mm 5000 L L 6mm M...

Page 7: ...ded to pre rinse the tip several times prior to use Do not pipette liquids with temperatures above 70 C NOTE When pipetting acids or corrosive solutions which emit vapors it is recommended to disassem...

Page 8: ...ISO 8655 guidelines for the maximum nominal liquid volume drawn by the pipettor and for 10 of the maximum or minimum liquid volume according to the values given Section 5 The pipettors are designed t...

Page 9: ...rm 10 aspirations and calculate the average value in mg Calculate the volume in L by multiplying the value in mg by the conversion factor Z L mg If the average aspirated volume exceeds the permissible...

Page 10: ...accuracy should not alter if the pipet ting processes including autoclaving are carried out as described in this manual If a change in accuracy occurs it is recommended to Check the calibration of the...

Page 11: ...ling elements are not sufficiently lubricated e g after repeated autoclaving procedures Incorrect aspiration Liquid with properties other than water den sity viscosity Calibrate the pipettor using the...