6 – VWR STandaRd Line PiPeTTORS
instruction Manual – 7
•
ambient operating temperature is +5°C to 45°C.
•
ambient storage conditions (in the original packaging
during transport and short storage) is -25°C to 55°C.
When working with the pipettor:
•
Follow general work safety regulations regarding
hazards related to work in the laboratory.
•
Take special care when pipetting aggressive
substances.
•
Use appropriate protective attire (e.g., clothing,
goggles, and gloves).
•
avoid pointing the pipettor at yourself or others
during use.
•
Only use parts and accessories recommended
by the manufacturer.
5. Specifications
The pipettor is a high quality instrument which offers
excellent accuracy and precision. The accuracy and
precision (repeatability) of the liquid volume depends on
the quality of pipet tips used. The values for accuracy
and precision shown in the table below were obtained
using manufacturer non-filter pipet tips. Those tips are
recommended for use to ensure compatibility, accuracy
and precision when pipetting.
Nominal
Volume
(μL)
Volume
(μL)
Accuracy
(%)
Precision
(%)
Non-filter
Tips (μL)
Single-channel Pipettors
2
Min.
0.1
±40.0
≤12.0
10
0.2
±12.0
≤6.0
1
±2.7
≤1.3
Max.
2
±1.5
≤0.7
10
Min.
0.5
±4.0
≤2.8
10
1.0
±2.5
≤1.8
5.0
±1.0
≤0.6
Max.
10.0
±0.5
≤0.4
20
Min.
2
±3.0
≤1.5
200
10
±1.0
≤0.5
Max.
20
±0.8
≤0.3
Nominal
Volume
(μL)
Volume
(μL)
Accuracy
(%)
Precision
(%)
Non-filter
Tips (μL)
100
Min.
10
±1.6
≤0.80
200
50
±0.8
≤0.24
Max.
100
±0.8
≤0.20
200
Min.
20
±1.2
≤0.60
200
100
±0.8
≤0.25
Max.
200
±0.6
≤0.20
1,000
Min.
100
±1.6
≤0.40
1,000
500
±0.7
≤0.20
Max. 1,000
±0.6
≤0.15
5,000
Min.
500
±1.2
≤0.50
5,000
1,000
±0.6
≤0.25
2,500
±0.6
≤0.20
Max. 5,000
±0.5
≤0.15
10,000
Min. 1,000
±2.5
≤0.6
10,000
5,000
±0.8
≤0.3
Max. 10,000
±0.5
≤0.2
8-channel and 12-channel Pipettors
8 - 10
12 - 10
Min.
0.5
±10.0
≤8.0
10
1
±8.0
≤6.0
5
±4.0
≤2.0
Max.
10
±2.0
≤1.2
8 - 50
12 - 50
Min.
5
±4.0
≤2.5
200
25
±3.0
≤1.2
Max.
50
±1.6
≤0.6
8 - 200
12 - 200
Min.
20
±3.0
≤1.5
200
100
±1.5
≤0.8
Max.
200
±1.0
≤0.6
8 - 300
12 - 300
Min.
50
±1.6
≤1.5
300
150
±1.2
≤1.0
Max.
300
±1.0
≤0.6
The accuracy and precision were obtained gravi-
metrically using manufacturer tips performing at least
10 measurements of distilled water at a temperature of
20°C ± 1°C according to en iSO 8655 standards.
The use of tips from other manufacturers or filter tips
may result in incorrect liquid aspiration and require
pipettor recalibration.