- 30 -
15.2. Probe number
The meter may store parameters of three probes, which may be replaced
without the need of entering the group again, only the number which
symbolises the sensor group has to be selected. By pressing the
PROBE NO.
window choose one of the three groups recorded under
E1
,
E2
or
E3
number.
15.3. Sensor group
The meter may cooperate with Pt1000B standard temperature sensor or with
Pt-1000S selected sensor of higher accuracy. Before starting measurement,
the probe data have to be entered.
To enter, press the
SENSOR GROUP
window, the numerical keyboard will
appear to insert the value given on the probe‘s plug (the number following the
G letter) and confirm with the
button.
The meter is calibrated and ready for the measurement.
To adjust the meter for cooperation with Pt-1000B standard sensor, into the
SENSOR GROUP
window insert the
nn00
number, where nn – cable lenght in
meters in range 1 - 19m. For the Pt1000B sensor on 1 meter cable the
number will be 100 (the zero at the frontal position is not displayed) and for
the same sensor on 15 meter cable the number will be 1500.
As standard the meter is set for co-operation with standard PT-1000B sensor.
15.4. Unit
The reading may be displayed in
0
C, K or
0
F.
By pressing the
UNIT
window choose the measurement unit.
15.5. Temperature of the manual compensation
After disconnecting the temperature probe the meter switches to manual
compensation automatically and calculates in the entered temperature value.
To enter this value, press the
MANUAL TEMP.
window, the numerical keyboard
will appear to enter the value and confirm with the
button.
User‘s manual for
PH-2500L
pH meter
v1.0x
Summary of Contents for 662-2380
Page 1: ...Instruction manual VWR pH METER PH 2500L EU cat no 662 2380 Version 1 1 Issue 12 2021 ...
Page 5: ... 1 I INTRODUCTION ...
Page 14: ... 10 II pH MEASUREMENT User s manual for PH 2500L pH meter v1 0x ...
Page 29: ... 25 III REDOX POTENTIAL AND TEMPERATURE MEASUREMENT ...
Page 36: ... 32 IV OPTIONS User s manual for PH 2500L pH meter v1 0x ...
Page 39: ... 35 Pic 14 ...