- 9 -
7.
PREPARATION TO WORK
Before starting measurements:
-
connect prepared combination pH electrode or redox electrode to the
F
(BNC-50) input;
-
connect the temperature probe to the
t
(Chinch/RCA) temperature input;
-
if work with power adapter is planned, connect it to the
P
input;
-
switch the meter on by pressing the
button.
7.1. Choosing the kind of the temperature compensation
The meter switches itself to the automatic or manual temperature
compensation mode. Connecting the temperature probe switches the
automatic temperature compensation on. Next to the reading the symbol is
displayed. After disconnecting the probe the meter enters the manual
temperature compensation mode. Instead of the symbol the is displayed.
In case of the manual temperature compensation the temperature value is
set on the temperature measurement parameters screen.