9
For DBL124 Binocular :
CAMERA -X3
To use Visicam software you have to tell the Software which is the Driver by selecting the microscope-shaped
icon.
GETTING STARTED WITH Visicam software, A VERY EASY MEASURING AND DOCUMENTATION
SOFTWARE
Run Visicam software by clicking on its icon on your desktop.
Select the right Driver for capturing the image. Note that after choosing the driver and before capturing the
image you can set the brightness, saturation, gamma ecc..
Then you can capture the image.
The image is now ready for being manipulated using Visicam software.
Troubleshooting
Review the information in the table below to troubleshoot operating problems.
PROBLEM
CAUSE
SOLUTION
LIGHT DOESN’T TURN ON
Power supply not connected Check that the 5Vdc power supply
jack is well inserted on the rear of the
microscope
Connected
Rotate the brightness adjustment
control and check if there is an
increase in the light output
IMAGE CANNOT BE SEEN OR
IS DARK.
The iris diaphragm aperture
is not completely opened
Completely open the iris diaphragm
aperture
Brightness level is low
Rotate the brightness adjustment
potentiometer
Objective is not aligned with
the optical axis
Rotate the nosepiece until an objective
is well inserted in the optical path (it
“clicks”)
IMAGE IS UNCLEAR, BLURRED
OR HAS INSUFFICIENT
CONTRAST.
Objectives or filters are dirty
Wipe them clean
The iris diaphragm aperture
is not opened correctly
Open the iris diaphragm aperture
completely
Condenser at wrong height
Hold the condenser and rotate it until
you see a uniform illumination