Installation, Operation, and Maintenance
VWR Zero Air Generator
Plumbing Connections
Connect
1/
4
” tubing from your source of compressed air to the
1
/
4
” NPT fitting on
the inlet filter. As you face the generator from the rear, the inlet filter is located on
the left (see Figure 1).
Note:
Make gas connections using PTFE tape on all threaded
fittings. Do not use other compounds to seal the threads,
as this can contaminate the purified air.
When making connections from the zero air generator to your piece of
equipment, be sure to use clean stainless steel or refrigeration grade copper
tubing. Do not use plastic tubing for the outlet line. Connect your
1
/
8
” outlet tubing
to the
1
/
4
” NPT fitting on the outlet filter. As you face the generator from the rear,
the outlet filter is located on the right for Model 26000-022 (see Figure 1 located
at the back of this bulletin). For Model 26000-020, the filter is on the left side
panel when facing the front of the generator (see Figure 2, located at the back of
this bulletin).
To ensure flow rates within the capacity of the generator, a flow meter should be
installed in the outlet line between the generator and the instrument using the
purified air.
4