VWR 2005 Installation & Operation Manual Download Page 6

 

 

OPERATION 

4.1 

Check power supply against unit serial plate.  They must match. 

4.2 

Plug service cord into the grounded electrical outlet. 

4.3 

Push the power switch to the I/On position, and turn the Overtemperature Thermostat to 
its maximum position, clockwise using a coin or flat edged tool. 

4.4 

Set  Main  Temperature  Controller:

  Enter  desired  set  point  temperature.    To  enter  set 

point  mode  on  the  controller,  press  either  the  Up  or  Down  arrow  pad  one  time.    The 
digital  display  will  start  to  blink,  going  from  bright  to  dim.    While  blinking,  the  digital 
display is showing the set point.  To change the set point, use the Up and Down arrow 
pads.    If  the  arrow  pads  are  not  pressed  for  five  (5)  seconds,  the  display  will  stop 
blinking  and  will  read  the  temperature  of  the  unit.    Note  that  the  Overtemperature 
Thermostat  should  be  turned  to  its  maximum  position,  (clockwise)  until  the  unit  has 
stabilized  at  desired  set  point  temperature.    Allow  the  incubator  at  least  24  hours  to 
stabilize. 

4.5 

Calibration:  

It is recommended that calibration is done once the unit is installed in its 

working  environment  and  has  been  stabile  at  set  point  for  several  hours.    Place  a 
certified  reference  thermometer  in  the  chamber.    Be  certain  the  thermometer  is  not 
touching  any  shelving.    Allow  the  temperature  to  stabilize  again  until  the  thermometer 
reads  a  constant  value  for  one  hour.    Compare  the  digital  display  with  the  reference 
thermometer.    If  there  is  an  unacceptable  difference,  put  the  display  into  calibration 
mode  by  pressing  both  the  Up  and  Down  arrow  pads  at  the  same  time  until  the  two 
outside decimal points begin to flash.  While the decimal points are flashing the display 
can  be  calibrated  by  pressing  the  Up  or  Down  arrow  pads  until  the  display  reads  the 
correct  value.    Allow  the  incubator  temperature  to  stabilize  again,  and  recalibrate  if 
necessary. 

4.6 

Set  Overtemperature  Thermostat:

    As  mentioned  previously,  the  Overtemperature 

Thermostat should be initially set to its maximum position, (clockwise), to allow the unit 
to  stabilize.    Once  the  incubator  is  stabile  at  the  desired  set  point,  turn  the 
Overtemperature  Thermostat  counterclockwise  until  the  OVERTEMPERATURE 
ACTIVATED light turns on.  Next, turn the Thermostat clockwise just until the light turns 
off.  Then  turn  the  Thermostat  clockwise  1/16  of  an  inch  past  the  point  where  the  light 
went  out.  This  will  set  the  Overtemperature  Thermostat  at  approximately  1

C  above 

main Temperature set point. 

4.7 

Accessory  Outlet:

    This  unit  has  been  supplied  with  a  1  amp  accessory  outlet  inside 

the chamber to provide power for equipment such as magnetic stirrers, rockers, etc.  Be 
certain the apparatus draws 1 amp or less.  The power switch on the front panel controls 
power  to  the  outlet.    Note  that  the  apparatus  may  provide  additional  heat  that  could 
affect the temperature range of this incubator.  It is recommended that testing be done 
with  the  incubator  and  any  accessory  equipment  to  insure  that  the  desired  operating 
conditions can be met. 

Section 

Summary of Contents for 2005

Page 1: ...LOW TEMPERATURE INCUBATOR MODEL 2005 INSTALLATION AND OPERATIONAL MANUAL Revised 1 2015 4861388 ...

Page 2: ... 7 0 PARTS LIST UNIT SPECIFICATIONS SCHEMATICS This unit is a general purpose Biochemical Oxygen Demand BOD incubator for professional industrial or educational use where the preparation or testing of materials is done at approximately atmospheric pressure and no flammable volatile or combustible materials are being heated This unit is not intended for hazardous or household locations or use ...

Page 3: ...to substantiate your claim 1 3 Return Shipment Save the shipping crate until you are sure all is well If for any reason you must return the unit first contact your customer representative for authorization Supply nameplate data including model number and serial number Please see the manual cover for information on where to contact Customer service 1 4 Accessories Verify that all of the equipment i...

Page 4: ...re air circulation around the unit allow a minimum of 20cm between incubator rear and sides and any walls partitions or obstructions to free airflow 2 3 Lifting Handling This unit is heavy and care should be taken to use appropriate lifting devices that are sufficiently rated for these loads Units should only be lifted from their bottom surfaces Doors handles and knobs are not adequate for lifting...

Page 5: ...roller It acts as an override in the event that the Main control fails in the ON position 3 5 Overtemperature Light This pilot lamp is marked OVERTEMPERATURE ACTIVATED and is on when the thermostat has been activated Under normal operating conditions this pilot light should never be on 3 6 Circuit Breaker Non CE units Adjacent to the power cord the circuit breaker is an added measure of protection...

Page 6: ...reference thermometer If there is an unacceptable difference put the display into calibration mode by pressing both the Up and Down arrow pads at the same time until the two outside decimal points begin to flash While the decimal points are flashing the display can be calibrated by pressing the Up or Down arrow pads until the display reads the correct value Allow the incubator temperature to stabi...

Page 7: ...7 4 8 Load the chamber with sample bottles See Figure 2 for bottle placement Figure 2 ...

Page 8: ...atile or flammable cleaners are evaporates and dry before reattaching the unit to the power supply 5 2 Compressor Compartments Located at the back and bottom of the unit the compressor compartment can collect dust that will inhibit proper air flow This compartment should be vacuumed out at least once every six 6 months to ensure maximum efficiency 5 3 Defrosting Generally defrosting should not be ...

Page 9: ...1 Verify that controller is asking for heat by looking for controller light 2 Do all controller functions work 3 Is the Overtemperature set high enough Indicated chamber temperature unstable 1 0 1 may be normal 2 Is fan working 3 Is ambient room temperature radically changing either door opening or room airflow from heaters or air conditioning stabilize ambient conditions 4 Overtemperature set too...

Page 10: ...ing its housing Adjust the motor mounting bracket position to re center the fan or blower wheel if necessary 2 Check the fan or blower wheel for damage or out of balance condition Replace the fan or blower wheel if it is damaged or out of balance 3 Turn the motor shaft to make sure that it spins freely If it binds or the bearings make a rubbing or scrapping sound then replace the motor Door not se...

Page 11: ... Factory Set Over Temperature Control 1750715 1750715 Heating Element 2350508 44 6 ohms 2350508 150 1 ohms I O On Off Power Switch 7850570 7850570 Main Temperature Controller 1750967 1750968 OTP and Compressor High Cut Thermostat 1750862 1750813 Pilot Light Green 4650554 4650554 Pilot Light Red 4650553 4650553 Power Cord USA 1800529 1800537 Step Down Transformer 8350515 Section 7 ...

Page 12: ...FICATIONS Weight Shipping Net 2005 130 lbs 108 lbs Dimensions Exterior WxDxH in Interior WxDxH in 2005 20 75 x 21 25 x 38 75 15 x 11 x 25 Capacity 3 cubic feet Temperature Range Uniformity 2005 0 to 45 C 0 5 C 20 C ...

Page 13: ...54 4880564 COMPRESSOR FAN 4880545 INTERNAL FAN THE COMPRESSOR IS PART OF THE BASE UNIT 1750862 PRESET TO 35 C 1750862 9570863 ELEMENT ASSY 1750715 155 F 68 C 2350508 ELEMENT COIL 45 SAFETY ACTIVATED HEATING ACTIVATED OUTLETS 6100525 POWER CORD AND PLUG 1100505 CIRCUIT BREAKER 15A GREEN LIGHTED POWER SWITCH 7850570 HOT NEUTRAL GROUND LOAD SENSOR TEMPERATURE CONTROL 1750967 ORG ...

Page 14: ... PART OF THE BASE UNIT 1750862 PRESET TO 35 C 1750813 9570864 ELEMENT ASSY 1750715 155 F 68 C 2350508 ELEMENT COIL SAFETY ACTIVATED HEATING ACTIVATED 2 4 1 1750813 BLK BLK BRN BRN 8350515 TRANSFORMER 240 120VAC 750VA YEL 2800502 EMI FILTER HOT NEUTRAL GROUND LOAD POWER CORD 1800537 BLK BLK BLK RED RED RED HT FUSE FUSE OUTLETS 6100526 GREEN LIGHTED POWER SWITCH 7850570 TEMPERATURE CONTROL 1750968 ...

Page 15: ...TX Lake Charles LA Mid Atlantic Area Baltimore MD Branford CT Bridgeport NJ S Plainfield NJ VWR Canlab Offices Mississauga Ontario Ville Mont Royal Québec Edmonton Alberta Or Call Direct for Specialized Service Locations VWR International Switzerland Ruchligstrasse 20 P O Box 464 Dietikon Switzerland CH 8953 011 41 1 745 1155 fax 011 41 1 745 1150 VWR Direct 911 Commerce Ct Buffalo Grove IL 60089 ...

Reviews: