VWR 10ECEEVWA Installation And Operations Download Page 15

GP, Explosion-Proof and FMS Lab Refrigerators and Freezers

Installation    |    11

How to Seal Killark® Box 
Conduit

To Seal Killark® Box Conduit with Fiber and Sealing 
Compound to Help Protect Against Explosions (Explosion-
Proof Units Only):

The purpose of the procedure that follows is to build fiber 
rope dams on the left and right hubs of the horizontal conduit. 
The fiber rope damns will surround conduit wiring that is 
housed inside the horizontal conduit. When both the left and 
right fiber rope dams have been pressed into place, sealing 
compound is poured between the two and forms into an 
airtight plug.

All of this is done in order to prevent the very real threat of gas 
entering the Killark box and a resulting serious explosion. After 
the unit wires have been pulled through the horizontal conduit 
the following procedure is required:

Turn power off at the circuit breaker before proceeding.

Place a small amount of sealing compound granules, 
enclosed, into a clean mixing vessel. Add small amounts 
of water while stirring until a thick paste is formed, then 
carefully continue adding smaller amounts of water until a 
thick gravy consistency is achieved—NOT WATERY. 
Discard any material that becomes too stiff to use. Never 
attempt to restore workability by stirring in more water.

Locate silver Killark box, back/top-center of unit.

Unscrew conduit domed-cover.

Unscrew and Remove Cover

Conduit

Horizontal

Left Hub

Right Hub

Conduit

Opening

Conduit Domed-Cover

Note: 

KILLARK® is a registered trademark of Hubbell 

Incorporated, Shelton, CT, USA.

Figure 1. Sealing the Killark Box

Note: 

Wires must be kept separated as shown in the 

Figure 2

Left Hub

Right Hub

Wiring

Figure 2. Horizontal Conduit, Cutaway

Insert fiber rope material down into horizontal conduit 
opening. Pressing down firmly, work the material into the 
left hub and—most importantly—being sure the material 
COMPLETELY SURROUNDS THE WIRING, from the top 
to the bottom, completely blocking this end of the 
horizontal conduit.

Figure 3. Fiber rope material in left hub

Insert fiber rope material down into horizontal conduit 
opening. Pressing down firmly, work the material into the 
right hub and—most importantly—being sure the 
material COMPLETELY SURROUNDS THE WIRING, from 
the top to the bottom, completely blocking this end of the 
horizontal conduit.

Figure 4. Fiber rope material in right hub

Summary of Contents for 10ECEEVWA

Page 1: ...VWR General Purpose Explosion Proof and Flammable Materials Storage Laboratory Refrigerators and Freezers Installation and Operation 057 658 00 Rev C February 2019...

Page 2: ...tice VWR makes no representations or warranties with respect to this manual In no event shall VWR be held liable for any damages direct or incidental arising from or related to the use of this manual...

Page 3: ...pose Refrigerator Freezer 7 Unpacking and Inspection 8 Shipping Carton 8 Visible Loss or Damage 8 Concealed Loss or Damage 8 Operating Standards 9 Installation 10 Selecting a Location 10 Leveling the...

Page 4: ...o Save Energy 13 Safety Tips 13 Maintenance 14 Cleaning the Unit 14 Cleaning Interior Exterior Door Gaskets 14 Cleaning the Condenser 14 Troubleshooting 15 Wiring Diagram 16 End of Life Care 17 Warran...

Page 5: ...Models 1 Models The table below shows the units covered in this operation and installation manual by model number Table 1 Applicable Models Refrigerators and Freezers Explosion Proof 10ECEEVWA Flamma...

Page 6: ...al Code unless otherwise noted Alert Signals ON OFF Safety Alert Important operating instructions To reduce the risk of injury or poor performance of the unit Read the user manual before putting the e...

Page 7: ...Resistant Corrosion Resistant does not mean Corrosion Proof Care in storage is still required Store only corrosive reagents samples which truly need reduced temperature storage Flammable corrosive mat...

Page 8: ...ess steel refrigeration components that are more resistant to attack from corrosive vapors than are aluminum or epoxy coated aluminum It is important to note that Explosion Proof models must be connec...

Page 9: ...rior notice Explosion Proof units require being hard wired into the building electrical system and therefore require a qualified electrician to install as per the local electrical requirements and NEC...

Page 10: ...h helping to lower operating costs The interior is made of porcelain enameled steel or ABS plastic VWR Scientific offers the largest selection of FMS units All models are delivered with 3 wire line co...

Page 11: ...color 10LCEEVWA Refrigerator Chamber Dimensions H x W x D inches cm 38 25 x 19 75 x 18 25 97 2 x 50 2 x 46 4 cm Freezer Chamber Dimensions H x W x D inches cm 18 25 x 18 x 11 25 46 3 x 45 7 x 28 6 cm...

Page 12: ...o adequately describe such external evidence of loss or damage may result in the carrier s refusing to honor your damage claim The form required to file such a claim will be supplied by the carrier IM...

Page 13: ...ment in a Pollution Degree 21 and Over voltage Category II environment These units are designed to operate under the following environmental conditions Indoor use Altitude up to 2000 m 6512 feet Maxim...

Page 14: ...al to that specified on the data plate WARNING For personal safety this unit must be properly grounded before use Flammable Materials Storage and General Purpose units must be connected to a grounded...

Page 15: ...tinue adding smaller amounts of water until a thick gravy consistency is achieved NOT WATERY Discard any material that becomes too stiff to use Never attempt to restore workability by stirring in more...

Page 16: ...ll occur in approximately 30 minutes However the sealing compound requires a minimum of 8 hours above 32 F 0 C to develop sufficient strength to withstand explosions Note Wires must be kept separated...

Page 17: ...eft at this setting for the nighttime period return temperature control to original setting when the room thermostat is returned to its normal setting Restart Procedure If unit is unplugged or turned...

Page 18: ...ld soap detergent and lukewarm water can be used for cleaning the interior exterior and door gaskets with a soft clean cloth Rinse with clean water and dry thoroughly before reconnecting and turning o...

Page 19: ...ure over 109 F 43 C Door opening frequency is high Defrost unit Clean the condenser with a dry brush or vacuum A leak around the door gasket will allow cold air to escape This causes unit to work hard...

Page 20: ...am GP Explosion Proof and FMS Lab Refrigerators and Freezers Wiring Diagram THERMOSTAT BLK WHT 120V BLK 240V OVERLOAD PROTECTOR RED C M S M S I C RELAY COMPRESSOR Figure 6 Wiring diagram for 10FCEEVWA...

Page 21: ...nit Some suggestions are listed below 1 Remove items and defrost unit Be sure to clean up any biological safety hazards 2 Remove the cabinet door to help prevent entrapment inside of a unit 3 Have a c...

Page 22: ...electrical power surges iv improper storage and handling of the products v use of the products in combination with equipment or software not supplied by VWR or vi installation maintenance repair serv...

Page 23: ...a 91 22 6716 2200 China 800 810 5118 400 650 5118 Japan 81 3 5826 1616 Australia 61 39757 4300 Austria 43 1 801 40 0 Belgium 32 53 73 42 41 France 33 2 2803 2180 Germany international 49 6184 90 6000...

Page 24: ...lding One Suite 200 100 Matsonford Road Radnor PA P O Box 6660 United States www vwr com Find out more at www vwr com 2019 VWR Inc All rights reserved All trademarks are the property of VWR and its su...

Reviews: