pH value
MU 6100 H
36
ba77077e02
09/2014
6.2.6
Calibration data
Displays the calibra-
tion data
The calibration record of the last calibration is to be found under the menu item,
<OK>
/
Calibration
/
Calibration record
. To open it, press the
<CAL
_
>
key in
the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e.g. to a PC, with the
<F2>
/
[USB output]
key.
Displaying the cali-
bration data memory
The calibration records of the last calibrations are to be found in the menu,
<OK>
/
Calibration
/
Calibration data storage
.
Calibration evalua-
tion
After calibrating, the meter automatically evaluates the calibration. The zero
point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
The calibration record is automatically transmitted to the interface
after calibrating.
Menu item
Setting/
function
Explanation
Calibration
/
Calibra-
tion data storage
/
Display
-
Displays the calibration record.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibra-
tion record to the interface with
<F2>
/
[USB output]
.
Quit the display with
<F1>
/
[Back]
or
<OK>
.
Switch directly to the mea-
sured value display with
<MODE>
.
Calibration
/
Calibra-
tion data storage
/
Output to USB
-
Outputs the calibration records to
the interface.