VWR International 89231-656 User Manual Download Page 6

2.

Attach the storage tube for the dissolved oxygen probe to the dissolved oxygen probe on the
multisensor.

3.

Connect the multisensor to the meter. Turn the meter on.

4.

Wait for the dissolved oxygen probe to polarize. Refer to 

Table 1

 for the amount of time.

5.

Enter the atmospheric pressure and salinity on the meter.

6.

Turn the storage tube for the dissolved oxygen probe and remove it from the dissolved oxygen
probe on the multisensor.

7.

Rinse the storage tube for the dissolved oxygen probe with deionized water.

8.

Invert and shake the storage tube for the dissolved oxygen probe to remove any water droplets.

Table 1  Polarization times

Disconnect time

Polarization time

Less than 5 minutes

10 minutes

5–15 minutes

45 minutes

More than 15 minutes

6 hours

Calibration

Before calibration:

Prepare the probe for use. Refer to 

Preparation for use

 on page 5.

Prepare the meter. Refer to the meter manual.

Calibration notes

• For maximum accuracy, calibrate the probe at least once a day.
• Prepare fresh standard solutions for the calibration.
• The calibration and sample measurement conditions must be as similar as possible (e.g., the

temperature of the solution, stir procedure, stir rate and position of the probe).

• For conductivity calibration, use 1-point calibration and select the standard nearest to the expected

measurement range.

• Calibration of conductivity cells is important as the conductivity cell constant can change with time

and calibration identifies the actual cell constant versus the nominal value.

• Do not dilute conductivity standards or samples.
• For dissolved oxygen calibration, a water-saturated air calibration is recommended. Refer to 

Dissolved oxygen probe calibration procedure – water-saturated (100%) air

 on page 7.

• When the probe is submerged, make sure that there are no air bubbles under the probe sensor

tip(s). Gently shake the probe from side to side to remove any air bubbles.

• Make sure that the reference junction is fully in the solution.
• Do not put the probe on the bottom or sides of the container.
• If stabilization is slow, shake the probe from side to side in the solution.
• If a calibration error occurs, refer to 

Troubleshooting

 on page 12.

6

   

English

Summary of Contents for 89231-656

Page 1: ...DOC022 53 80319 sympHony Multisensor Probes 07 2012 Edition 1 User Manual...

Page 2: ......

Page 3: ...te For return for recycling please contact the equipment producer or supplier for instructions on how to return end of life equipment producer supplied electrical accessories and all auxiliary items f...

Page 4: ...or use with sympHony handheld meters only Product overview The multisensor probes are combination gel filled electrodes for pH conductivity and ORP or dissolved oxygen measurements in general aqueous...

Page 5: ...e equipment to remove broken glass To prepare the probe for calibration or sample measurement 1 Turn the probe storage tube s and remove from the probe Keep the probe storage tube s 2 Rinse the probe...

Page 6: ...pare fresh standard solutions for the calibration The calibration and sample measurement conditions must be as similar as possible e g the temperature of the solution stir procedure stir rate and posi...

Page 7: ...onstant rate to prevent the formation of a vortex 6 Push Read Stir gently The display shows Stable when the reading is stable 7 When the reading is stable the meter prompts for the next calibration po...

Page 8: ...dissolved oxygen calibration polarize the probe Refer to Polarize the dissolved oxygen probe 89231 660 multisensor on page 5 2 Push Calibrate and select the probe if applicable 3 Prepare the standard...

Page 9: ...page 5 Calibrate the probe Refer to Calibration on page 6 The manufacturer recommends that the probe is calibrated at least once a day for the best measurement accuracy Measurement notes The calibrati...

Page 10: ...surements are done prepare the probe for storage Refer to Storage on page 12 Maintenance Clean the probe N O T I C E Do not rub or touch the tip of the pH or dissolved oxygen sensors Clean the probe w...

Page 11: ...fer to Fill the membrane module on page 11 Replace the membrane module N O T I C E Carefully handle the membrane modules to prevent damage It is necessary to replace the membrane module periodically R...

Page 12: ...ge solution to the mark Refer to Specifications on page 3 for the applicable storage solution 2 Put the probe s in the storage tube s and turn to tighten Rehydrate the probe conductivity and pH sensor...

Page 13: ...ectrode SAME BUFFERS Calibrate the probe Refer to Calibration on page 6 Examine the probe Refer to Examine the probe on page 13 Examine the standard solution Use a new standard solution Table 4 Measur...

Page 14: ...14 English...

Page 15: ......

Page 16: ...alid in USA void where prohibited by law or company policy while supplies last Visit vwr com to view our policy and additional disclaimers VWR forms of VWR and the VWR logo and or design are either re...

Reviews: