When calibration is performed, the time and date of calibration is recorded, making it important to
ensure that the EX1200A
’s time is configured properly. To determine the time, click on the
Time
Configuration
link in the command menu. If the time is displayed incorrectly, use the following
steps to configure the time from the
Time Configuration
web page.
1)
Select the correct
Time Zone
from the drop menu.
2)
Set the
Time Source
to
Manual
.
3)
Enable the
Set Time
checkbox.
4)
Enter the date in the
Date (MM:DD:YYYY)
fields.
5)
Enter the time in time in the
Time (HH:MM:SS)
fields using a 24-hour format.
6)
Click the
Submit
button.
F
IGURE
9-5:
T
IME CONFIGURATION
P
ERFORMING
C
ALIBRATION
Once connected to the EX1200A and after ensuring that the time is set properly, it can be
calibrated using the SFP. The following description provides describes how this is done. Prior to
beginning calibration, ensure the following:
The EX1200A DMM (device under calibration), the function generator, the DMM, and the
calibrator/amplifier must all be powered up for a minimum of one hour before calibration
The Agilent
3458A “Auto Cal All” procedure must be performed
prior to calibration.
The 5700A’s “Cal Zero” procedure must be performed
prior to calibration.
Ensure that
the “I FUSE” is installed in the
EX1200A DMM.
NOTE
Make sure that temporary files and java cache are removed before staring the calibration, See
Remove Temporary Internet Files
3) Select
Set Time
5) Set
Time
(24 HR format)
4) Set
Date
2) Set to
Manual
6) Click when done
1) Select
None
Summary of Contents for Ametek EX1200A Series
Page 33: ......
Page 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Page 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Page 65: ......
Page 73: ......
Page 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Page 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Page 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Page 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Page 155: ......
Page 161: ......