21
4.4
Operating procedure
4.4.1
Measurement process
Connect the measured signals according to the selected operating mode (see
section 4.5 of this Operating manual).
Connection is carried out by coaxial cables with wave resistance of 50 Ohm and
SMA type connectors. The tightening torque of SMA connectors should be between
0.8 and 1.1 N·m.
The device will be ready for use with the guaranteed frequency instability given
in Table 2 after a warm-up period.
Start the measurements in the Multichannel frequency comparator program
following the instructions in the "Multichannel frequency comparator VCH-315M
User guide".
You can view the instability characteristics of the frequency of measured
sinusoidal signals during or after the end of the measurements following the
instructions in the "Multichannel frequency comparator VCH-315M User guide" and
"Program for processing measurement data of frequency comparators/counters"
Analyzer "User guide".