III. Protocol: Setup Instructions - GO
2
VENT™ (continued)
4.3 HAZARDS/PRECAUTIONS
4.3.1 The GO
2
VENT™ should be used only by individuals who have
adequate training in CPR techniques and the operation of gas-
powered resuscitators.
4.3.2 Do not use grease or oil on the GO
2
VENT™ for any reason.
4.3.3 Do not use the GO
2
VENT™™ in oxygen deficient atmospheres
or near open flames.
4.3.4 Do not smoke while using the GO
2
VENT™ or any other oxygen
equipment.
4.3.5 Do not dismantle or attempt to remove any components other
than those required for routine operations. Any tampering with
the GO
2
VENT™ may cause the unit to malfunction, and will
automatically void the warranty.
4.4 SET-UP INSTRUCTIONS
4.4.1 The GO
2
VENT™ is suitable for patients weighing 22 pounds or
10 kilograms.
4.4.2 Select desired FiO2 delivery.
[a] If 100% FiO
2
is
to be delivered to
the patient, connect
tubing to the white
gas connector with
the DISS thread
connection on the
patient tee. Ensure
that the green knob
is turned clockwise
until it comes to a
stop.
[b] If 50% FiO2 is
to be delivered to
the patient,
connect tubing to
the white gas
connector with the
DISS thread
connection on the
patient tee. Ensure
that the green knob
is turned
counterclockwise
until it comes to a
stop.
ENTRAINED
FLOWCHART
100%
50%