41
a) DC voltage (V
)
Range
Accuracy
Resolution
4.000 V
±(0.8 % r8 digits)
1 mV
40.00 V
10 mV
400.0 V
100 mV
600 V
1 V
Overload protection: 600 V
Input impedance: approx. 10 MΩ
b) AC voltage (V~)
Range
Accuracy
Resolution
400.0 mV
±(2.0 % r10 digits)
0.1 mV
4.000 V
±(1.6 % r4 digits)
1 mV
40.00 V
10 mV
400.0 V
100 mV
600 V
1 V
Overload protection: 600 V
Input impedance: approx. 4.5 MΩ
Frequency response: 45 – 400 Hz
Display: Sine wave valid value (average value response)
c) DC current (A
)
Range
Accuracy
Resolution
µA
400 µA
±(1.3 % r3 digits)
0.1 µA
4000 µA
1 µA
mA
40.00 mA
±(1.6 % r2 digits)
10 µA
400.0 mA
0.1 mA
A
4.000 A
±(2.0 % r10 digits)
1 mA
10.0 A
10 mA
Overload protection:
mAμA range: F1 fuse 6 x 32 mm F 0.5 A H 600 V (CE) Item no.: 702632
10A range: F2 fuse 6 x 25 mm F 10 A H 600 V (CE) Item no.: 700161
Measuring operations of >5 A may only be performed for at most 10 seconds with a subsequent
measuring pause of 15 minutes.