background image

9 Operation

9.1 Distance and precision

Fig. 1
To obtain accurate measurements, consider the following points when measuring:

 

The measured object must be larger than the IR measuring spot 

(fig. 1).

 

The measured object should be at least twice the size of the IR measuring spot 

S

 (fig. 1).

 

The smaller the object, the closer it must be to the thermometer.

 

The D:S ratio is 6:1. At a distance 

(fig. 1) of 6 cm, the size of the IR measuring spot is 

1 cm.

9.2 Surface and environmental constraints

Not all surface materials are suitable for taking measurements. Observe the following con

-

straints:

 

Do not measure surface temperatures of shiny or polished metal surfaces.

 

Do not measure through transparent surfaces such as glass. It will measure the surface 

temperature of the glass instead.

 

Steam, dust, and smoke can affect the accuracy.

9.3 Switching on/off

1. 

Press the measuring button to switch the product on.

 

Æ

The display lights up.

 

Æ

The product turns off automatically after approx. 15 seconds of inactivity.

9.4 Enabling/disabling the target laser

The target laser helps you direct the infrared sensor at the desired location.

 

 CAUTION

Eye injury from laserlight

 

 

Do not point the laser beam at people or animals.

1.  Press the measuring button to switch on.

2. 

Press the button 

 to enable/disable the target laser.

 

Æ

(If enabled) the symbol 

 is shown on the display.

 

Æ

(If enabled) the target laser is active while the measuring button is pressed.

9.5 Measuring temperature

Important:

Long measurements of high temperatures at close range cause the product to self-heat. 

This results in inaccurate measurements.
The higher the temperature, the greater the measuring distance and the shorter the meas

-

uring duration should be.

1

Precondition

 

D

The product has adjusted to the ambient temperature. This can take up to 30 min.

2

Procedure

1. 

To measure, press and hold the measuring button and point the infrared sensor at the 

object.

 

Æ

The measured temperature and 

SCAN

 are shown on the display.

 

Æ

If the temperature lies outside the measuring range, the display shows 

----

.

2.  Release the measuring button to freeze the measurement.

 

Æ

The last measured value and 

HOLD

 appear on the display.

9.6 Measuring differential temperature

With  the  differential  temperature  function  you  measure  the  difference  between  the  highest 

and the lowest temperature reading relative to the temperature of a reference object. The first 

measurement serves as the reference temperature.

Temperature

Measuring

 

location

Reference 

temperature

Max

Min

Fig. 2
The graph shows how temperature can vary across measuring locations (or 

measuring objects) in relation to the initially set reference temperature.

1.  Press the measuring button to switch on.

2.  Repeatedly press the button 

MODE

 until 

DIF

 is shown on the display.

3. 

Point the infrared sensor at the object.

4. 

Press and hold the measuring button to start taking measurements.

 

Æ

The display shows the current temperature and the temperature difference 

ΔT

 between 

the highest and the lowest temperature reading (fig. 2).

9.7 Displaying Maximum/Minimum/Average readings

Besides temperature, the product records and calculates maximum, minimum, and average 

temperature. You can set which reading is displayed when taking measurements.

1.  Press the measuring button to switch on.

2.  Repeatedly press the button 

MODE

 to select minimum (

MIN

), maximum (

MAX

), or average 

(

AVG

) temperature.

9.8 Setting temperature alarms

The temperature alarm function allows you to set a lower and upper temperature threshold. 

When measurements fall below the lower or exceed the upper threshold, an alarm sounds. 

Thresholds can be enabled independently of each other.

1

Setting thresholds

1.  Press and hold the button 

MODE

 until 

ε=

 flashes on the display.

2.  Repeatedly press the button 

MODE

 until 

 (upper threshold) or 

 (lower threshold)  

flashes on the display.

3. 

Press the button ▲to activate the threshold.

 

Æ

The display switches from 

Off

 (deactivated) to 

On

 (activated).

4.  Press the button 

MODE

 to confirm.

 

Æ

The temperature flashes on the display.

5. 

Use the buttons ▲ and ▼to set the threshold temperature.

6.  Continue pressing the button 

MODE

 until no symbols flash on the display.

 

Æ

The corresponding threshold symbol shows on the display.

 

Æ

You have set and activated the threshold.

2

Disabling thresholds

Refer to the procedure above for setting thresholds. Instead of activating them, you deactivate 

them. 

10 

Configuration

Use the configuration menu to change and adjust display and measurement settings.

10.1 Available settings

Setting

Display indicator

Description

Temperature unit

°C/°F 

Select between degrees Celsius or degrees 

Fahrenheit.

Emissivity

ε=

Set the emissivity level.
Many organic materials have an emissivity of 

0.95.
Refer to “Emissivity reference values” for 

emissivity levels of different materials.

10.2 Adjusting settings

1.  Press and hold the button 

MODE

 until 

ε=

 flashes on the display.

2.  Repeatedly press the button 

MODE

 until the display indicator corresponding to the setting 

you intend to change flashes on the display.

3. 

Use the buttons ▲ and ▼to adjust the setting values.

4.  Press the button 

MODE

 to save the changes.

5. 

Exit the menu by pressing the button 

MODE

 repeatedly until no symbols flash on the dis

-

play.

 

Æ

You have changed the setting.

Reviews: