Measuring tolerances
Accuracy in ± (% of r display error in counts (= number of smallest points)). These accuracy readings are
valid for one year at a temperature of +23 °C (± 5 °C) and a relative humidity of less than 75 % (non-condensing). If
the multimeter is used outside of this temperature range, use the following coefficient to calculate the accuracy. +0.1 x
(specified accuracy)/1 °C
The accuracy of measurements may be affected when the multimeter is used in a high-frequency electromagnetic
field.
Direct voltage (V/DC)
Range
Resolution
Accuracy
60.00 mV*
0.01 mV
±(1.2% + 8)
600.0 mV*
0.1 mV
±(1.0% + 8)
6.000 V
0.001 V
±(0.9% + 4)
60.00 V
0.01 V
600.0 V
0.1 V
*Only available in “mV” mode
Specified measurement range: 5 - 100% of the measurement range
600 V overload protection; Impedance: 10 MΩ (mV: ≤1000 MΩ)
The multimeter may display ≤10 counts if a measurement input is short-circuited.
Direct voltage (V/DC) LoZ
Range
Resolution
Accuracy
6.000 V
0.001 V
±(1.8% + 7)
60.00 V
0.01 V
600.0 V
0.1 V
Specified measurement range: 5 - 100% of the measurement range
600 V overload protection; Impedance: 400 kΩ (max. 250 V, 3 secs)
The multimeter may display ≤10 counts if a measurement input is short-circuited.
After using the LoZ feature, leave the multimeter for 1 minute before using it again.